CVE-2025-1204
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP addres…
Mitigation
- Vendor workaround: Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
- If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
Who should check
- Anyone running Contec Health CMS8000 Patient Monitor (versions: see the vendor advisory)
Vulnerability facts
- Affected
- Contec Health CMS8000 Patient Monitor
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, no login needed
- CVSS
- 7.7 (v4.0, CNA)CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:NOpen in CVSS calculator
Show all 11 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Attack requirements: Present
- Privileges required: None
- User interaction: Passive
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- Confidentiality impact on other systems: None
- Integrity impact on other systems: None
- Availability impact on other systems: None
- EPSS
- 0.77% chance of exploitation in 30 days (percentile 52, 2026-11-26)
- CWE
- CWE-912
- Modified
Description
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Contec Health CMS8000 Patient Monitor | — | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2025-1204. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2025-1204.md Affected: Contec Health CMS8000 Patient Monitor Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Contec Health CMS8000 Patient Monitor: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2025-1204 being exploited?
There is no confirmed exploitation: CVE-2025-1204 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.77% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2025-1204?
High: CVSS 7.7 (v4.0, scored by CNA). Remote, no login needed. A user has to take an action, such as opening a link, for the attack to work.
Which versions are affected by CVE-2025-1204?
Contec Health CMS8000 Patient Monitor.
Is there a fix for CVE-2025-1204?
No fixed version is listed in the sources yet. Vendor workaround: Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
How do I check if I'm affected by CVE-2025-1204?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.