CVE-2025-0626
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
Mitigation
- Vendor workaround: Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
- If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
- Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .
Who should check
- Anyone running Contec Health CMS8000 Patient Monitor All versions
Vulnerability facts
- Affected
- Contec Health CMS8000 Patient Monitor All versions
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, no login needed
- CVSS
- 7.7 (v4.0, CNA)CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:NOpen in CVSS calculator
Show all 11 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Attack requirements: Present
- Privileges required: None
- User interaction: Passive
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- Confidentiality impact on other systems: None
- Integrity impact on other systems: None
- Availability impact on other systems: None
- EPSS
- 0.95% chance of exploitation in 30 days (percentile 60, 2026-11-26)
- CWE
- CWE-912
- Modified
Description
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Contec Health CMS8000 Patient Monitor | All versions | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2025-0626. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2025-0626.md Affected: Contec Health CMS8000 Patient Monitor All versions Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Contec Health CMS8000 Patient Monitor: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
Other
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
- https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
- https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
FAQ
Is CVE-2025-0626 being exploited?
There is no confirmed exploitation: CVE-2025-0626 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.95% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2025-0626?
High: CVSS 7.7 (v4.0, scored by CNA). Remote, no login needed. A user has to take an action, such as opening a link, for the attack to work.
Which versions are affected by CVE-2025-0626?
Contec Health CMS8000 Patient Monitor All versions.
Is there a fix for CVE-2025-0626?
No fixed version is listed in the sources yet. Vendor workaround: Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks. If asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120. Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .