CVE-2019-25719
Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running Dräger Infinity Acute Care System <= VG4.1.1; <= VG4.0.3
- Anyone running Dräger Standalone Infinity M540 patient monitor <= VG4.1.1
Vulnerability facts
- Affected
- Dräger Infinity Acute Care System <= VG4.1.1; <= VG4.0.3; Dräger Standalone Infinity M540 patient monitor <= VG4.1.1
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, no login needed
- CVSS
- 8.8 (v4.0, CNA)CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:NOpen in CVSS calculator
Show all 11 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Attack requirements: None
- Privileges required: None
- User interaction: None
- Confidentiality impact: Low
- Integrity impact: Low
- Availability impact: High
- Confidentiality impact on other systems: None
- Integrity impact on other systems: None
- Availability impact on other systems: None
- EPSS
- 0.13% chance of exploitation in 30 days (percentile 1, 2026-11-26)
- CWE
- CWE-924
- Modified
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Affected and fixed versions
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2019-25719. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2019-25719.md Affected: Dräger Infinity Acute Care System <= VG4.1.1; <= VG4.0.3; Dräger Standalone Infinity M540 patient monitor <= VG4.1.1 Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Dräger Infinity Acute Care System, Dräger Standalone Infinity M540 patient monitor: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
Vendor advisories
FAQ
Is CVE-2019-25719 being exploited?
There is no confirmed exploitation: CVE-2019-25719 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.13% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2019-25719?
High: CVSS 8.8 (v4.0, scored by CNA). Remote, no login needed.
Which versions are affected by CVE-2019-25719?
Dräger Infinity Acute Care System <= VG4.1.1; <= VG4.0.3; Dräger Standalone Infinity M540 patient monitor <= VG4.1.1.
Is there a fix for CVE-2019-25719?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.