# CVE-2019-25719: Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** HIGH 8.8
- **CVSS:** 8.8 (v4.0, CNA) `CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N`
- **CVSS metrics:** Attack vector: Network; Attack complexity: Low; Attack requirements: None; Privileges required: None; User interaction: None; Confidentiality impact: Low; Integrity impact: Low; Availability impact: High; Confidentiality impact on other systems: None; Integrity impact on other systems: None; Availability impact on other systems: None
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Remote, no login needed
- **EPSS:** 0.13% (percentile 1, 2026-11-26)
- **CWE:** CWE-924
- **Published:** 2026-06-02T14:00:52.578Z
- **Modified:** 2026-07-22T19:10:00.120Z

## Who should check

- Anyone running Dräger Infinity Acute Care System \<= VG4.1.1; \<= VG4.0.3
- Anyone running Dräger Standalone Infinity M540 patient monitor \<= VG4.1.1

## Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

## Affected and fixed versions

- Dräger Infinity Acute Care System: \<= VG4.1.1; \<= VG4.0.3 (cvelist)
- Dräger Standalone Infinity M540 patient monitor: \<= VG4.1.1 (cvelist)

## Mitigation

- No fix listed yet: follow the vendor advisory.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2019-25719. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2019-25719.md
Affected: Dräger Infinity Acute Care System <= VG4.1.1; <= VG4.0.3; Dräger Standalone Infinity M540 patient monitor <= VG4.1.1
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Dräger Infinity Acute Care System, Dräger Standalone Infinity M540 patient monitor: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2019-25719 being exploited?

There is no confirmed exploitation: CVE-2019-25719 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.13% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2019-25719?

High: CVSS 8.8 (v4.0, scored by CNA). Remote, no login needed.

### Which versions are affected by CVE-2019-25719?

Dräger Infinity Acute Care System \<= VG4.1.1; \<= VG4.0.3; Dräger Standalone Infinity M540 patient monitor \<= VG4.1.1.

### Is there a fix for CVE-2019-25719?

No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.

### How do I check if I'm affected by CVE-2019-25719?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- vendor: <https://static.draeger.com/security/download/PSA-19-255-02_Product-Security-Advisory-IACS-VG4.1.pdf>
- other: <https://www.vulncheck.com/advisories/dr-ger-infinity-m540-vg4-spoofing-and-dos-via-network-message-handling>

## Attribution

- CVE Program: CVE® is a registered trademark of The MITRE Corporation. CVE records © 1999–2026 The MITRE Corporation. <https://www.cve.org/Legal/TermsOfUse>
- NVD: This product uses data from the NVD API but is not endorsed or certified by the NVD. <https://nvd.nist.gov/vuln/detail/CVE-2019-25719>

HTML: <https://flawdb.com/CVE-2019-25719> · JSON: <https://flawdb.com/api/v1/entries/CVE-2019-25719>