CVE-2026-17583
Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check
Mitigation
- Vendor workaround: Applied Biosystems 3130 Series Data Collection Software: Product is End of Life (EoL), no update provided
- ABI PRISM 3100/3100-Avant Data Collection Software: Product is End of Life (EoL), no update provided
- ABI PRISM 310 Data Collection Software: Product is End of Life (EoL), no update provided
- For users who are unable to immediately implement all applicable security updates, Thermo Fisher Scientific recommends implementing the following interim mitigation measures until the applicable updates have been installed:
- * Maintain a secure chain of custody for files generated by the HID instrumentation throughout the analysis workflow.
- * Store generated files on encrypted, password-protected storage media (for example, encrypted USB drives or encrypted hard drives).
- * Restrict access to generated files to authorized personnel in accordance with your laboratory's access control policies.
- * Apply the principle of least privilege by limiting user permissions on systems operating the HID instrumentation or hosting associated data analysis and secondary analysis software.
- * Leverage firewall rules and network access control lists (NACLs) to restrict internet connectivity to only trusted sources.
- For more information, refer to Thermo Fisher's security bulletin.
- https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/fsa_hid_bulletin.pdf
Who should check
- Anyone running Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software <= 4.0.2
- Anyone running Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software <= 5.0.2
- Anyone running Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software <= 1.2.5
- Anyone running Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software <= 1.2.0
- Anyone running Thermo Fisher Applied Biosystems GeneMapper ID-X Software <= 1.7.3
- Anyone running Thermo Fisher Applied Biosystems 3130 Series Data Collection Software <= 4.1
- Anyone running Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software <= 2.0
- Anyone running Thermo Fisher ABI PRISM 310 Data Collection Software <= 3.1
Vulnerability facts
- Affected
- Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software <= 4.0.2; Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software <= 5.0.2; Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software <= 1.2.5; Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software <= 1.2.0; Thermo Fisher Applied Biosystems GeneMapper ID-X Software <= 1.7.3; Thermo Fisher Applied Biosystems 3130 Series Data Collection Software <= 4.1; Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software <= 2.0; Thermo Fisher ABI PRISM 310 Data Collection Software <= 3.1
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 8.3 (v4.0, CNA)CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:H/SA:NOpen in CVSS calculator
Show all 11 CVSS metrics
- Attack vector: Local
- Attack complexity: Low
- Attack requirements: None
- Privileges required: None
- User interaction: None
- Confidentiality impact: None
- Integrity impact: High
- Availability impact: High
- Confidentiality impact on other systems: None
- Integrity impact on other systems: High
- Availability impact on other systems: None
- EPSS
- 0.14% chance of exploitation in 30 days (percentile 2, 2026-11-26)
- CWE
- CWE-353
- Modified
Description
The affected Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software | <= 4.0.2 | — | cvelist |
| Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software | <= 5.0.2 | — | cvelist |
| Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software | <= 1.2.5 | — | cvelist |
| Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software | <= 1.2.0 | — | cvelist |
| Thermo Fisher Applied Biosystems GeneMapper ID-X Software | <= 1.7.3 | — | cvelist |
| Thermo Fisher Applied Biosystems 3130 Series Data Collection Software | <= 4.1 | — | cvelist |
| Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software | <= 2.0 | — | cvelist |
| Thermo Fisher ABI PRISM 310 Data Collection Software | <= 3.1 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2026-17583. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2026-17583.md Affected: Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software <= 4.0.2; Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software <= 5.0.2; Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software <= 1.2.5; Thermo Fisher Applied Biosyste… Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software, Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software, Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software, Thermo Fisher Applied Biosystems SeqStudio Flex Series In…: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
Patches
FAQ
Is CVE-2026-17583 being exploited?
There is no confirmed exploitation: CVE-2026-17583 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.14% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2026-17583?
High: CVSS 8.3 (v4.0, scored by CNA). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2026-17583?
Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software <= 4.0.2; Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software <= 5.0.2; Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software <= 1.2.5; Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software <= 1.2.0; Thermo Fisher Applied Biosystems GeneMapper ID-X Software <= 1.7.3; Thermo Fisher Applied Biosystems 3130 Series Data Collection Software <= 4.1; Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software <= 2.0; Thermo Fisher ABI PRISM 310 Data Collection Software <= 3.1.
Is there a fix for CVE-2026-17583?
No fixed version is listed in the sources yet. Vendor workaround: Applied Biosystems 3130 Series Data Collection Software: Product is End of Life (EoL), no update provided ABI PRISM 3100/3100-Avant Data Collection Software: Product is End of Life (EoL), no update provided ABI PRISM 310 Data Collection Software: Product is End of Life (EoL), no update provided For users who are unable to immediately implement all applicable security updates, Thermo Fisher Scientific recommends implementing the following interim mitigation measures until the applicable updates have been installed: * Maintain a secure chain of custody for files generated by the HID instrumentation throughout the analysis workflow. * Store generated files on encrypted, password-protected storage media (for example, encrypted USB drives or encrypted hard drives). * Restrict access to generated files to authorized personnel in accordance with your laboratory's access control policies. * Apply the principle of least privilege by limiting user permissions on systems operating the HID instrumentation or hosting associated data analysis and secondary analysis software. * Leverage firewall rules and network access control lists (NACLs) to restrict internet connectivity to only trusted sources. For more information, refer to Thermo Fisher's security bulletin. https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/fsa_hid_bulletin.pdf