CVE-2025-20060
Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Application Exposure of Private Personal Information to an Unauthorized Actor
Mitigation
- Upgrade USB-C Blood Glucose Monitoring System Starter Kit Android Applications to 5.8.7.0.36 or later.
- Vendor workaround: Dario Health recommends users perform the following mitigations:
- * Update the application from trusted sources.
- * Don't use rooted/jailbroken devices.
- * Avoid public untrusted network.
- * For more information contact Dario Health https://www.dariohealth.com/contact/ directly.
Who should check
- Anyone running Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications < 5.8.7.0.36
- Anyone running Dario Health Dario Application Database and Internet-based Server Infrastructure All versions
Vulnerability facts
- Affected
- Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications < 5.8.7.0.36; Dario Health Dario Application Database and Internet-based Server Infrastructure All versions
- Fixed in
- Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications 5.8.7.0.36
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, no login needed
- CVSS
- 8.7 (v4.0, CNA)CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:NOpen in CVSS calculator
Show all 11 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Attack requirements: None
- Privileges required: None
- User interaction: None
- Confidentiality impact: High
- Integrity impact: None
- Availability impact: None
- Confidentiality impact on other systems: None
- Integrity impact on other systems: None
- Availability impact on other systems: None
- EPSS
- 0.71% chance of exploitation in 30 days (percentile 49, 2026-11-26)
- CWE
- CWE-359
- Modified
Description
An attacker could expose cross-user personal identifiable information (PII) and personal health information transmitted to the Android device via the Dario Health application database.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications | < 5.8.7.0.36 | — | cvelist |
| Dario Health Dario Application Database and Internet-based Server Infrastructure | All versions | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2025-20060. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2025-20060.md Affected: Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications < 5.8.7.0.36; Dario Health Dario Application Database and Internet-based Server Infrastructure All versions Fixed in: Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications 5.8.7.0.36 Steps: 1. Search this machine and project for Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications, Dario Health Dario Application Database and Internet-based Server Infrastructure: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2025-20060 being exploited?
There is no confirmed exploitation: CVE-2025-20060 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.71% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2025-20060?
High: CVSS 8.7 (v4.0, scored by CNA). Remote, no login needed.
Which versions are affected by CVE-2025-20060?
Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications < 5.8.7.0.36; Dario Health Dario Application Database and Internet-based Server Infrastructure All versions.
Is there a fix for CVE-2025-20060?
Yes. Fixed in Dario Health USB-C Blood Glucose Monitoring System Starter Kit Android Applications 5.8.7.0.36. Upgrade USB-C Blood Glucose Monitoring System Starter Kit Android Applications to 5.8.7.0.36 or later. Vendor workaround: Dario Health recommends users perform the following mitigations: * Update the application from trusted sources. * Don't use rooted/jailbroken devices. * Avoid public untrusted network. * For more information contact Dario Health https://www.dariohealth.com/contact/ directly.
How do I check if I'm affected by CVE-2025-20060?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.