CVE-2025-1863
Insecure default settings for recorder products
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running Yokogawa Electric Corporation GX10 / GX20 / GP10 / GP20 Paperless Recorders R5.04.01 or earlier
- Anyone running Yokogawa Electric Corporation GM Data Acquisition System R5.05.01 or earlier
- Anyone running Yokogawa Electric Corporation DX1000 / DX2000 / DX1000N Paperless Recorders R4.21 or earlier
- Anyone running Yokogawa Electric Corporation FX1000 Paperless Recorders R1.31 or earlier
- Anyone running Yokogawa Electric Corporation μR10000 / μR20000 Chart Recorders R1.51 or earlier
- Anyone running Yokogawa Electric Corporation MW100 Data Acquisition Units All versions
- Anyone running Yokogawa Electric Corporation DX1000T / DX2000T Paperless Recorders All versions
- Anyone running Yokogawa Electric Corporation CX1000 / CX2000 Paperless Recorders All versions
Vulnerability facts
- Affected
- Yokogawa Electric Corporation GX10 / GX20 / GP10 / GP20 Paperless Recorders R5.04.01 or earlier; Yokogawa Electric Corporation GM Data Acquisition System R5.05.01 or earlier; Yokogawa Electric Corporation DX1000 / DX2000 / DX1000N Paperless Recorders R4.21 or earlier; Yokogawa Electric Corporation FX1000 Paperless Recorders R1.31 or earlier; Yokogawa Electric Corporation μR10000 / μR20000 Chart Recorders R1.51 or earlier; Yokogawa Electric Corporation MW100 Data Acquisition Units All versions; Yokogawa Electric Corporation DX1000T / DX2000T Paperless Recorders All versions; Yokogawa Electric Corporation CX1000 / CX2000 Paperless Recorders All versions
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, no login needed
- CVSS
- 9.8 (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- EPSS
- 0.96% chance of exploitation in 30 days (percentile 60, 2026-11-26)
- CWE
- CWE-1188
- Modified
Description
Insecure default settings have been found in recorder products provided by Yokogawa Electric Corporation. The default setting of the authentication function is disabled on the affected products. Therefore, when connected to a network with default settings, anyone can access all functions related to settings and operations. As a result, an attacker can illegally manipulate and configure important data such as measured values and settings. This issue affects GX10 / GX20 / GP10 / GP20 Paperless Recorders: R5.04.01 or earlier; GM Data Acquisition System: R5.05.01 or earlier; DX1000 / DX2000 / DX1000N Paperless Recorders: R4.21 or earlier; FX1000 Paperless Recorders: R1.31 or earlier; μR10000 / μR20000 Chart Recorders: R1.51 or earlier; MW100 Data Acquisition Units: All versions; DX1000T / DX2000T Paperless Recorders: All versions; CX1000 / CX2000 Paperless Recorders: All versions.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Yokogawa Electric Corporation GX10 / GX20 / GP10 / GP20 Paperless Recorders | R5.04.01 or earlier | — | cvelist |
| Yokogawa Electric Corporation GM Data Acquisition System | R5.05.01 or earlier | — | cvelist |
| Yokogawa Electric Corporation DX1000 / DX2000 / DX1000N Paperless Recorders | R4.21 or earlier | — | cvelist |
| Yokogawa Electric Corporation FX1000 Paperless Recorders | R1.31 or earlier | — | cvelist |
| Yokogawa Electric Corporation μR10000 / μR20000 Chart Recorders | R1.51 or earlier | — | cvelist |
| Yokogawa Electric Corporation MW100 Data Acquisition Units | All versions | — | cvelist |
| Yokogawa Electric Corporation DX1000T / DX2000T Paperless Recorders | All versions | — | cvelist |
| Yokogawa Electric Corporation CX1000 / CX2000 Paperless Recorders | All versions | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2025-1863. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2025-1863.md Affected: Yokogawa Electric Corporation GX10 / GX20 / GP10 / GP20 Paperless Recorders R5.04.01 or earlier; Yokogawa Electric Corporation GM Data Acquisition System R5.05.01 or earlier; Yokogawa Electric Corporation DX1000 / DX2000 / DX1000N Paperless Recorders R4.21 or earlier; Yokogawa Electric Corporation … Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Yokogawa Electric Corporation GX10 / GX20 / GP10 / GP20 Paperless Recorders, Yokogawa Electric Corporation GM Data Acquisition System, Yokogawa Electric Corporation DX1000 / DX2000 / DX1000N Paperless Recorders, Yokogawa Electric Corporation FX1000 Paperless Recorders, Yokogawa Electric Corporation…: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2025-1863 being exploited?
There is no confirmed exploitation: CVE-2025-1863 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.96% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2025-1863?
Critical: CVSS 9.8 (v3.1, scored by CNA). Remote, no login needed.
Which versions are affected by CVE-2025-1863?
Yokogawa Electric Corporation GX10 / GX20 / GP10 / GP20 Paperless Recorders R5.04.01 or earlier; Yokogawa Electric Corporation GM Data Acquisition System R5.05.01 or earlier; Yokogawa Electric Corporation DX1000 / DX2000 / DX1000N Paperless Recorders R4.21 or earlier; Yokogawa Electric Corporation FX1000 Paperless Recorders R1.31 or earlier; Yokogawa Electric Corporation μR10000 / μR20000 Chart Recorders R1.51 or earlier; Yokogawa Electric Corporation MW100 Data Acquisition Units All versions; Yokogawa Electric Corporation DX1000T / DX2000T Paperless Recorders All versions; Yokogawa Electric Corporation CX1000 / CX2000 Paperless Recorders All versions.
Is there a fix for CVE-2025-1863?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2025-1863?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.