CVE-2025-15653
Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
Mitigation
- Upgrade Zeus IE to 1.0.5 or later.
- Upgrade Zeus RS C500 to 1.0.9 or later.
Who should check
- Anyone running Dräger Zeus IE < 1.0.5
- Anyone running Dräger Zeus RS C500 < 1.0.9
Vulnerability facts
- Affected
- Dräger Zeus IE < 1.0.5; Dräger Zeus RS C500 < 1.0.9
- Fixed in
- Dräger Zeus IE 1.0.5, Dräger Zeus RS C500 1.0.9
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 7.0 (v4.0, CNA)CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:NOpen in CVSS calculator
Show all 11 CVSS metrics
- Attack vector: Physical
- Attack complexity: Low
- Attack requirements: None
- Privileges required: None
- User interaction: None
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- Confidentiality impact on other systems: None
- Integrity impact on other systems: None
- Availability impact on other systems: None
- EPSS
- 0.26% chance of exploitation in 30 days (percentile 11, 2026-11-26)
- CWE
- CWE-668
- Modified
Description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Affected and fixed versions
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2025-15653. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2025-15653.md Affected: Dräger Zeus IE < 1.0.5; Dräger Zeus RS C500 < 1.0.9 Fixed in: Dräger Zeus IE 1.0.5, Dräger Zeus RS C500 1.0.9 Steps: 1. Search this machine and project for Dräger Zeus IE, Dräger Zeus RS C500: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
Vendor advisories
FAQ
Is CVE-2025-15653 being exploited?
There is no confirmed exploitation: CVE-2025-15653 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.26% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2025-15653?
High: CVSS 7.0 (v4.0, scored by CNA). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2025-15653?
Dräger Zeus IE < 1.0.5; Dräger Zeus RS C500 < 1.0.9.
Is there a fix for CVE-2025-15653?
Yes. Fixed in Dräger Zeus IE 1.0.5, Dräger Zeus RS C500 1.0.9. Upgrade Zeus IE to 1.0.5 or later. Upgrade Zeus RS C500 to 1.0.9 or later.
How do I check if I'm affected by CVE-2025-15653?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.