Skip to content
FlawDB
Sign in
Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

CVE-2025-14097

HIGH 7.2EPSS 59%

Remote Code Execution Vulnerability in Radiometer Products

Mitigation

  • Vendor workaround: If the network is not considered secure, please remove the analyzer from the network.

Who should check

  • Anyone running Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems
  • Anyone running Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems
  • Anyone running Radiometer Medical Aps AQT90 FLEX Analyzers Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems

Vulnerability facts

Affected
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems; Radiometer Medical Aps AQT90 FLEX Analyzers Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems
Fixed in
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers 3.5MR11, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20MR2
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Remote, login needed
CVSS
7.2 (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Network
  • Attack complexity: Low
  • Privileges required: High
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: High
  • Integrity impact: High
  • Availability impact: High
EPSS
0.93% chance of exploitation in 30 days (percentile 59, 2026-11-26)
CWE
CWE-287
Published
Modified

Description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Affected and fixed versions

Package / productAffectedFixedSource
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS AnalyzersApplication software versions < 3.5MR11 with Windows 7, Windows XP operating systems—cvelist
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20MR2 with Windows 7, Windows XP operating systems—cvelist
Radiometer Medical Aps AQT90 FLEX AnalyzersApplication software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2025-14097. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2025-14097.md
Affected: Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems; R…
Fixed in: Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers 3.5MR11, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20MR2

Steps:
1. Search this machine and project for Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers, Radiometer Medical Aps AQT90 FLEX Analyzers: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2025-14097 being exploited?

There is no confirmed exploitation: CVE-2025-14097 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.93% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2025-14097?

High: CVSS 7.2 (v3.1, scored by CNA). Remote, login needed.

Which versions are affected by CVE-2025-14097?

Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems; Radiometer Medical Aps AQT90 FLEX Analyzers Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems.

Is there a fix for CVE-2025-14097?

Yes. Fixed in Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers 3.5MR11, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20MR2. Vendor workaround: If the network is not considered secure, please remove the analyzer from the network.

How do I check if I'm affected by CVE-2025-14097?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors