Skip to content
FlawDB
Sign in
Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

CVE-2025-14096

HIGH 8.4EPSS 8%

Credential Disclosure vulnerability in Radiometer Products

Mitigation

  • Vendor workaround: Only authorized people can physically access the analyzer.

Who should check

  • Anyone running Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system
  • Anyone running Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system
  • Anyone running Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system; Windows XP Operating system

Vulnerability facts

Affected
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system; Windows XP Operating system
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
8.4 (v3.1, CNA)CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Local
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: High
  • Integrity impact: High
  • Availability impact: High
EPSS
0.22% chance of exploitation in 30 days (percentile 8, 2026-11-26)
CWE
CWE-798, CWE-250
Published
Modified

Description

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Affected and fixed versions

Package / productAffectedFixedSource
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS AnalyzersWindows 7 Operating system; Windows XP Operating system—cvelist
Radiometer Medical Aps AQT90 FLEX AnalyzersWindows 7 Operating system; Windows XP Operating system—cvelist
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzersWindows 7 Operating system; Windows XP Operating system—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2025-14096. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2025-14096.md
Affected: Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Opera…
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers, Radiometer Medical Aps AQT90 FLEX Analyzers, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2025-14096 being exploited?

There is no confirmed exploitation: CVE-2025-14096 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.22% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2025-14096?

High: CVSS 8.4 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

Which versions are affected by CVE-2025-14096?

Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system; Windows XP Operating system.

Is there a fix for CVE-2025-14096?

No fixed version is listed in the sources yet. Vendor workaround: Only authorized people can physically access the analyzer.

How do I check if I'm affected by CVE-2025-14096?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors