CVE-2025-14096
Credential Disclosure vulnerability in Radiometer Products
Mitigation
- Vendor workaround: Only authorized people can physically access the analyzer.
Who should check
- Anyone running Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system
- Anyone running Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system
- Anyone running Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system; Windows XP Operating system
Vulnerability facts
- Affected
- Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system; Windows XP Operating system
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 8.4 (v3.1, CNA)CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Local
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- EPSS
- 0.22% chance of exploitation in 30 days (percentile 8, 2026-11-26)
- CWE
- CWE-798, CWE-250
- Modified
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers | Windows 7 Operating system; Windows XP Operating system | — | cvelist |
| Radiometer Medical Aps AQT90 FLEX Analyzers | Windows 7 Operating system; Windows XP Operating system | — | cvelist |
| Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers | Windows 7 Operating system; Windows XP Operating system | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2025-14096. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2025-14096.md Affected: Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Opera… Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers, Radiometer Medical Aps AQT90 FLEX Analyzers, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2025-14096 being exploited?
There is no confirmed exploitation: CVE-2025-14096 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.22% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2025-14096?
High: CVSS 8.4 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2025-14096?
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps AQT90 FLEX Analyzers Windows 7 Operating system; Windows XP Operating system; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers Windows 7 Operating system; Windows XP Operating system.
Is there a fix for CVE-2025-14096?
No fixed version is listed in the sources yet. Vendor workaround: Only authorized people can physically access the analyzer.
How do I check if I'm affected by CVE-2025-14096?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.