Privilege boundary violation in Radiometer Products

Mitigation

  • Upgrade ABL9 Analyzers to 1.5.0 or later.
  • Vendor workaround: Only authorized people can physically access the analyzer.

Who should check

  • Anyone running Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers All application software versions with Windows 7, Windows XP as underlying OS; Application software versions < 3.5MR11 with Windows 10 as underlying OS
  • Anyone running Radiometer Medical Aps AQT90 FLEX Analyzers All Application software versions <= 8.13 MR2
  • Anyone running Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS; Application software versions < 6.20 MR2 with Windows 10 as underlying OS
  • Anyone running Radiometer Medical Aps ABL9 Analyzers Application software versions < 1.5.0

Vulnerability facts

Affected
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers All application software versions with Windows 7, Windows XP as underlying OS; Application software versions < 3.5MR11 with Windows 10 as underlying OS; Radiometer Medical Aps AQT90 FLEX Analyzers All Application software versions <= 8.13 MR2; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS; Application software versions < 6.20 MR2 with Windows 10 as underlying OS; Radiometer Medical Aps ABL9 Analyzers Application software versions < 1.5.0
Fixed in
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers 3.5MR11, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20, Radiometer Medical Aps ABL9 Analyzers 1.5.0
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
6.8 (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Physical
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: High
  • Integrity impact: High
  • Availability impact: High
EPSS
0.28% chance of exploitation in 30 days (percentile 13, 2026-11-26)
CWE
CWE-284, CWE-693
Published
Modified

Description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Affected and fixed versions

Package / productAffectedFixedSource
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS AnalyzersAll application software versions with Windows 7, Windows XP as underlying OS; Application software versions < 3.5MR11 with Windows 10 as underlying OS—cvelist
Radiometer Medical Aps AQT90 FLEX AnalyzersAll Application software versions <= 8.13 MR2—cvelist
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX AnalyzersApplication software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS; Application software versions < 6.20 MR2 with Windows 10 as underlying OS—cvelist
Radiometer Medical Aps ABL9 AnalyzersApplication software versions < 1.5.0—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2025-14095. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2025-14095.md
Affected: Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers All application software versions with Windows 7, Windows XP as underlying OS; Application software versions < 3.5MR11 with Windows 10 as underlying OS; Radiometer Medical Aps AQT90 FLEX Analyzers All Application software versions <= 8…
Fixed in: Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers 3.5MR11, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20, Radiometer Medical Aps ABL9 Analyzers 1.5.0

Steps:
1. Search this machine and project for Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers, Radiometer Medical Aps AQT90 FLEX Analyzers, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers, Radiometer Medical Aps ABL9 Analyzers: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2025-14095 being exploited?

There is no confirmed exploitation: CVE-2025-14095 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.28% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2025-14095?

Medium: CVSS 6.8 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

Which versions are affected by CVE-2025-14095?

Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers All application software versions with Windows 7, Windows XP as underlying OS; Application software versions < 3.5MR11 with Windows 10 as underlying OS; Radiometer Medical Aps AQT90 FLEX Analyzers All Application software versions <= 8.13 MR2; Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OS; Application software versions < 6.20 MR2 with Windows 10 as underlying OS; Radiometer Medical Aps ABL9 Analyzers Application software versions < 1.5.0.

Is there a fix for CVE-2025-14095?

Yes. Fixed in Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers 3.5MR11, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20, Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers 6.20, Radiometer Medical Aps ABL9 Analyzers 1.5.0. Upgrade ABL9 Analyzers to 1.5.0 or later. Vendor workaround: Only authorized people can physically access the analyzer.

How do I check if I'm affected by CVE-2025-14095?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.