CVE-2024-48973
Debug port on Life2000 Ventilator serial interface is enabled by default
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running Baxter Life2000 Ventilation System 06.08.00.00 and prior
Vulnerability facts
- Affected
- Baxter Life2000 Ventilation System 06.08.00.00 and prior
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 9.3 (v3.1, CNA)CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Local
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Changed (impact reaches other components)
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- EPSS
- 0.27% chance of exploitation in 30 days (percentile 12, 2026-11-26)
- CWE
- CWE-1263
- Modified
Description
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Baxter Life2000 Ventilation System | 06.08.00.00 and prior | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2024-48973. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2024-48973.md Affected: Baxter Life2000 Ventilation System 06.08.00.00 and prior Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Baxter Life2000 Ventilation System: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2024-48973 being exploited?
There is no confirmed exploitation: CVE-2024-48973 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.27% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2024-48973?
Critical: CVSS 9.3 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2024-48973?
Baxter Life2000 Ventilation System 06.08.00.00 and prior.
Is there a fix for CVE-2024-48973?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2024-48973?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.