CVE-2024-27107
Weak account password in GE HealthCare EchoPAC products
Mitigation
- Upgrade EchoPAC Software Only to 206.82 or later.
Who should check
- Anyone running GE HealthCare EchoPAC Software Only < 206.82
- Anyone running GE HealthCare ImageVault (versions: see the vendor advisory)
- Anyone running GE HealthCare EchoPAC Turnkey (versions: see the vendor advisory)
Vulnerability facts
- Affected
- GE HealthCare EchoPAC Software Only < 206.82; GE HealthCare ImageVault; GE HealthCare EchoPAC Turnkey
- Fixed in
- GE HealthCare EchoPAC Software Only 206.82
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 9.6 (v3.1, CNA)CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Adjacent network
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Changed (impact reaches other components)
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- EPSS
- 0.42% chance of exploitation in 30 days (percentile 28, 2026-11-26)
- CWE
- CWE-798
- Modified
Description
Weak account password in GE HealthCare EchoPAC products
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| GE HealthCare EchoPAC Software Only | < 206.82 | — | cvelist |
| GE HealthCare ImageVault | — | — | cvelist |
| GE HealthCare EchoPAC Turnkey | — | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2024-27107. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2024-27107.md Affected: GE HealthCare EchoPAC Software Only < 206.82; GE HealthCare ImageVault; GE HealthCare EchoPAC Turnkey Fixed in: GE HealthCare EchoPAC Software Only 206.82 Steps: 1. Search this machine and project for GE HealthCare EchoPAC Software Only, GE HealthCare ImageVault, GE HealthCare EchoPAC Turnkey: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2024-27107 being exploited?
There is no confirmed exploitation: CVE-2024-27107 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.42% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2024-27107?
Critical: CVSS 9.6 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2024-27107?
GE HealthCare EchoPAC Software Only < 206.82; GE HealthCare ImageVault; GE HealthCare EchoPAC Turnkey.
Is there a fix for CVE-2024-27107?
Yes. Fixed in GE HealthCare EchoPAC Software Only 206.82. Upgrade EchoPAC Software Only to 206.82 or later.
How do I check if I'm affected by CVE-2024-27107?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.