CVE-2023-1968

CRITICAL 10.0EPSS 57%

CVE-2023-1968

Mitigation

  • Vendor workaround: Illumina recommends using the UCS Vulnerability Instructions Guide https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html  based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software.

Who should check

  • Anyone running Illumina iScan Control Software 4.0.0; 4.0.5
  • Anyone running Illumina MiniSeq Control Software 2.0
  • Anyone running Illumina iSeq 100 All versions
  • Anyone running Illumina MiSeq Control Software 4.0 (RUO Mode)
  • Anyone running Illumina MiSeqDx Operating Software 4.0.1
  • Anyone running Illumina NextSeq 500/550 Control Software 4.0
  • Anyone running Illumina NextSeq 550Dx Control Software 4.0 (RUO Mode)
  • Anyone running Illumina NextSeq 550Dx Operating Software >= 1.0.0 , <= 1.3.1; 1.3.3

Vulnerability facts

Affected
Illumina iScan Control Software 4.0.0; 4.0.5; Illumina MiniSeq Control Software 2.0 ; Illumina iSeq 100 All versions; Illumina MiSeq Control Software 4.0 (RUO Mode); Illumina MiSeqDx Operating Software 4.0.1 ; Illumina NextSeq 500/550 Control Software 4.0; Illumina NextSeq 550Dx Control Software 4.0 (RUO Mode); Illumina NextSeq 550Dx Operating Software >= 1.0.0 , <= 1.3.1; 1.3.3 ; Illumina NextSeq 1000/2000 Control Software <= 1.4.1; Illumina NovaSeq 6000 Control Software <= 1.7 ; Illumina NovaSeq Control Software 1.8
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Remote, no login needed
CVSS
10.0 (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Network
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: None
  • Scope: Changed (impact reaches other components)
  • Confidentiality impact: High
  • Integrity impact: High
  • Availability impact: High
EPSS
0.88% chance of exploitation in 30 days (percentile 57, 2026-11-26)
CWE
CWE-1327
Published
Modified

Description

Instruments with Illumina Universal Copy Service v2.x are vulnerable due to binding to an unrestricted IP address. An unauthenticated malicious actor could use UCS to listen on all IP addresses, including those capable of accepting remote communications.

Affected and fixed versions

Package / productAffectedFixedSource
Illumina iScan Control Software4.0.0; 4.0.5—cvelist
Illumina MiniSeq Control Software2.0 —cvelist
Illumina iSeq 100All versions—cvelist
Illumina MiSeq Control Software4.0 (RUO Mode)—cvelist
Illumina MiSeqDx Operating Software4.0.1 —cvelist
Illumina NextSeq 500/550 Control Software4.0—cvelist
Illumina NextSeq 550Dx Control Software4.0 (RUO Mode)—cvelist
Illumina NextSeq 550Dx Operating Software>= 1.0.0 , <= 1.3.1; 1.3.3 —cvelist
Illumina NextSeq 1000/2000 Control Software<= 1.4.1—cvelist
Illumina NovaSeq 6000 Control Software<= 1.7 —cvelist
Illumina NovaSeq Control Software1.8—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2023-1968. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2023-1968.md
Affected: Illumina iScan Control Software 4.0.0; 4.0.5; Illumina MiniSeq Control Software 2.0 ; Illumina iSeq 100 All versions; Illumina MiSeq Control Software 4.0 (RUO Mode); Illumina MiSeqDx Operating Software 4.0.1 ; Illumina NextSeq 500/550 Control Software 4.0; Illumina NextSeq 550Dx Control Software 4.…
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Illumina iScan Control Software, Illumina MiniSeq Control Software, Illumina iSeq 100, Illumina MiSeq Control Software, Illumina MiSeqDx Operating Software, Illumina NextSeq 500/550 Control Software: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2023-1968 being exploited?

There is no confirmed exploitation: CVE-2023-1968 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.88% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2023-1968?

Critical: CVSS 10.0 (v3.1, scored by CNA). Remote, no login needed.

Which versions are affected by CVE-2023-1968?

Illumina iScan Control Software 4.0.0; 4.0.5; Illumina MiniSeq Control Software 2.0 ; Illumina iSeq 100 All versions; Illumina MiSeq Control Software 4.0 (RUO Mode); Illumina MiSeqDx Operating Software 4.0.1 ; Illumina NextSeq 500/550 Control Software 4.0; Illumina NextSeq 550Dx Control Software 4.0 (RUO Mode); Illumina NextSeq 550Dx Operating Software >= 1.0.0 , <= 1.3.1; 1.3.3 ; Illumina NextSeq 1000/2000 Control Software <= 1.4.1; Illumina NovaSeq 6000 Control Software <= 1.7 ; Illumina NovaSeq Control Software 1.8.

Is there a fix for CVE-2023-1968?

No fixed version is listed in the sources yet. Vendor workaround: Illumina recommends using the UCS Vulnerability Instructions Guide https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html  based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software.

How do I check if I'm affected by CVE-2023-1968?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors