# CVE-2023-1966: CVE-2023-1966

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** HIGH 7.4
- **CVSS:** 7.4 (v3.1, CNA) `CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H`
- **CVSS metrics:** Attack vector: Physical; Attack complexity: Low; Privileges required: Low; User interaction: None; Scope: Changed (impact reaches other components); Confidentiality impact: High; Integrity impact: High; Availability impact: High
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Not remotely exploitable (local or adjacent access needed)
- **EPSS:** 1.01% (percentile 62, 2026-11-26)
- **CWE:** CWE-250
- **Published:** 2023-04-28T18:06:51.431Z
- **Modified:** 2026-06-17T05:29:09.353Z

## Who should check

- Anyone running Illumina iScan Control Software 4.0.0; 4.0.5
- Anyone running Illumina MiniSeq Control Software 2.0 
- Anyone running Illumina iSeq 100 All versions
- Anyone running Illumina MiSeq Control Software 4.0 (RUO Mode)
- Anyone running Illumina MiSeqDx Operating Software 4.0.1 
- Anyone running Illumina NextSeq 500/550 Control Software 4.0
- Anyone running Illumina NextSeq 550Dx Control Software 4.0 (RUO Mode)
- Anyone running Illumina NextSeq 550Dx Operating Software \>= 1.0.0 , \<= 1.3.1; 1.3.3 

## Description

Instruments with Illumina Universal Copy Service v1.x and
v2.x contain an unnecessary privileges vulnerability. An unauthenticated
malicious actor could upload and execute code remotely at the operating system
level, which could allow an attacker to change settings, configurations,
software, or access sensitive data on the affected product.

## Affected and fixed versions

- Illumina iScan Control Software: 4.0.0; 4.0.5 (cvelist)
- Illumina MiniSeq Control Software: 2.0  (cvelist)
- Illumina iSeq 100: All versions (cvelist)
- Illumina MiSeq Control Software: 4.0 (RUO Mode) (cvelist)
- Illumina MiSeqDx Operating Software: 4.0.1  (cvelist)
- Illumina NextSeq 500/550 Control Software: 4.0 (cvelist)
- Illumina NextSeq 550Dx Control Software: 4.0 (RUO Mode) (cvelist)
- Illumina NextSeq 550Dx Operating Software: \>= 1.0.0 , \<= 1.3.1; 1.3.3  (cvelist)
- Illumina NextSeq 1000/2000 Control Software: \<= 1.4.1 (cvelist)
- Illumina NovaSeq 6000 Control Software: \<= 1.7  (cvelist)
- Illumina NovaSeq Control Software: 1.8 (cvelist)

## Mitigation

- Vendor workaround: Illumina recommends using the  UCS Vulnerability Instructions Guide https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html  based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2023-1966. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2023-1966.md
Affected: Illumina iScan Control Software 4.0.0; 4.0.5; Illumina MiniSeq Control Software 2.0 ; Illumina iSeq 100 All versions; Illumina MiSeq Control Software 4.0 (RUO Mode); Illumina MiSeqDx Operating Software 4.0.1 ; Illumina NextSeq 500/550 Control Software 4.0; Illumina NextSeq 550Dx Control Software 4.…
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Illumina iScan Control Software, Illumina MiniSeq Control Software, Illumina iSeq 100, Illumina MiSeq Control Software, Illumina MiSeqDx Operating Software, Illumina NextSeq 500/550 Control Software: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2023-1966 being exploited?

There is no confirmed exploitation: CVE-2023-1966 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 1.01% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2023-1966?

High: CVSS 7.4 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

### Which versions are affected by CVE-2023-1966?

Illumina iScan Control Software 4.0.0; 4.0.5; Illumina MiniSeq Control Software 2.0 ; Illumina iSeq 100 All versions; Illumina MiSeq Control Software 4.0 (RUO Mode); Illumina MiSeqDx Operating Software 4.0.1 ; Illumina NextSeq 500/550 Control Software 4.0; Illumina NextSeq 550Dx Control Software 4.0 (RUO Mode); Illumina NextSeq 550Dx Operating Software \>= 1.0.0 , \<= 1.3.1; 1.3.3 ; Illumina NextSeq 1000/2000 Control Software \<= 1.4.1; Illumina NovaSeq 6000 Control Software \<= 1.7 ; Illumina NovaSeq Control Software 1.8.

### Is there a fix for CVE-2023-1966?

No fixed version is listed in the sources yet. Vendor workaround: Illumina recommends using the  UCS Vulnerability Instructions Guide https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html  based on the user’s specific system configuration to mitigate the vulnerabilities. Illumina recommends users read the instructions before downloading any software.

### How do I check if I'm affected by CVE-2023-1966?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- vendor: <https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html>
- other: <https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01>

## Attribution

- CVE Program: CVE® is a registered trademark of The MITRE Corporation. CVE records © 1999–2026 The MITRE Corporation. <https://www.cve.org/Legal/TermsOfUse>
- NVD: This product uses data from the NVD API but is not endorsed or certified by the NVD. <https://nvd.nist.gov/vuln/detail/CVE-2023-1966>

HTML: <https://flawdb.com/CVE-2023-1966> · JSON: <https://flawdb.com/api/v1/entries/CVE-2023-1966>