# CVE-2022-43557: BD BodyGuard™ Pumps – RS-232 Interface Vulnerability

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** MEDIUM 5.3
- **CVSS:** 5.3 (v3.1, CNA) `CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H`
- **CVSS metrics:** Attack vector: Physical; Attack complexity: High; Privileges required: None; User interaction: None; Scope: Unchanged; Confidentiality impact: Low; Integrity impact: Low; Availability impact: High
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Not remotely exploitable (local or adjacent access needed)
- **EPSS:** 0.28% (percentile 13, 2026-11-26)
- **CWE:** CWE-1299
- **Published:** 2022-12-05T00:00:00.000Z
- **Modified:** 2026-06-17T05:06:50.423Z

## Who should check

- Anyone running Becton, Dickinson and Company (BD) BodyGuard™ Pump BD BodyGuard™ ; CME BodyGuard™ 323 (2nd Edition); CME BodyGuard™ 323 Color Vision (2nd Edition); CME BodyGuard™ 323 Color Vision (3rd Edition); CME BodyGuard™ Twins (2nd Edition)

## Description

The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.

## Affected and fixed versions

- Becton, Dickinson and Company (BD) BodyGuard™ Pump: BD BodyGuard™ ; CME BodyGuard™ 323 (2nd Edition); CME BodyGuard™ 323 Color Vision (2nd Edition); CME BodyGuard™ 323 Color Vision (3rd Edition); CME BodyGuard™ Twins (2nd Edition) (cvelist)

## Mitigation

- Vendor workaround: BD recommends the following mitigations and compensating controls to reduce risk associated with this
- vulnerability:
- • Ensure physical access controls are in place and only authorized end-users have access to
- BD BodyGuard™ pumps.
- • Ensure that only BD-approved equipment is connected to the RS-232 interface of the affected pumps.
- • Ensure that no equipment is connected to the RS-232 interface when the affected pumps are delivering
- infusions.
- • Protect connected computer systems with BodyComm™ software with standard security measures.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2022-43557. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2022-43557.md
Affected: Becton, Dickinson and Company (BD) BodyGuard™ Pump BD BodyGuard™ ; CME BodyGuard™ 323 (2nd Edition); CME BodyGuard™ 323 Color Vision (2nd Edition); CME BodyGuard™ 323 Color Vision (3rd Edition); CME BodyGuard™ Twins (2nd Edition)
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Becton, Dickinson and Company (BD) BodyGuard™ Pump: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2022-43557 being exploited?

There is no confirmed exploitation: CVE-2022-43557 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.28% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2022-43557?

Medium: CVSS 5.3 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). The attack needs conditions outside the attacker's control, which makes it harder.

### Which versions are affected by CVE-2022-43557?

Becton, Dickinson and Company (BD) BodyGuard™ Pump BD BodyGuard™ ; CME BodyGuard™ 323 (2nd Edition); CME BodyGuard™ 323 Color Vision (2nd Edition); CME BodyGuard™ 323 Color Vision (3rd Edition); CME BodyGuard™ Twins (2nd Edition).

### Is there a fix for CVE-2022-43557?

No fixed version is listed in the sources yet. Vendor workaround: BD recommends the following mitigations and compensating controls to reduce risk associated with this vulnerability: • Ensure physical access controls are in place and only authorized end-users have access to BD BodyGuard™ pumps. • Ensure that only BD-approved equipment is connected to the RS-232 interface of the affected pumps. • Ensure that no equipment is connected to the RS-232 interface when the affected pumps are delivering infusions. • Protect connected computer systems with BodyComm™ software with standard security measures.

### How do I check if I'm affected by CVE-2022-43557?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- vendor: <https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-bodyguard-pumps-rs-232-interface-vulnerability>

## Attribution

- CVE Program: CVE® is a registered trademark of The MITRE Corporation. CVE records © 1999–2026 The MITRE Corporation. <https://www.cve.org/Legal/TermsOfUse>
- NVD: This product uses data from the NVD API but is not endorsed or certified by the NVD. <https://nvd.nist.gov/vuln/detail/CVE-2022-43557>

HTML: <https://flawdb.com/CVE-2022-43557> · JSON: <https://flawdb.com/api/v1/entries/CVE-2022-43557>