Skip to content
FlawDB
Sign in
Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

CVE-2022-38069

MEDIUM 4.3EPSS 20%

Contec Health CMS8000

Mitigation

  • Vendor workaround: Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.
  • The following mitigations could assist in reducing the risk for exploitation of vulnerabilities:
  • Disabling UART functionality at the CPU level
  • Enforcing unique device authentication before granting access to the terminal / bootloader
  • Where possible, enforcing secure boot.
  • Tamper stickers on the device casing to indicate when a device has been opened

Who should check

  • Anyone running Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All

Vulnerability facts

Affected
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
4.3 (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:LOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Physical
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: Low
  • Integrity impact: Low
  • Availability impact: Low
EPSS
0.34% chance of exploitation in 30 days (percentile 20, 2026-11-26)
CWE
CWE-798
Published
Modified

Description

Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters

Affected and fixed versions

Package / productAffectedFixedSource
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient MonitorAll—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2022-38069. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2022-38069.md
Affected: Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2022-38069 being exploited?

There is no confirmed exploitation: CVE-2022-38069 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.34% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2022-38069?

Medium: CVSS 4.3 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

Which versions are affected by CVE-2022-38069?

Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All.

Is there a fix for CVE-2022-38069?

No fixed version is listed in the sources yet. Vendor workaround: Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened

How do I check if I'm affected by CVE-2022-38069?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors