CVE-2022-3027
Contec Health CMS8000
Mitigation
- Vendor workaround: Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.
- The following mitigations could assist in reducing the risk for exploitation of vulnerabilities:
- Disabling UART functionality at the CPU level
- Enforcing unique device authentication before granting access to the terminal / bootloader
- Where possible, enforcing secure boot.
- Tamper stickers on the device casing to indicate when a device has been opened
Who should check
- Anyone running Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All
Vulnerability facts
- Affected
- Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 5.7 (v3.1, CNA)CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Adjacent network
- Attack complexity: Low
- Privileges required: None
- User interaction: Required
- Scope: Unchanged
- Confidentiality impact: None
- Integrity impact: None
- Availability impact: High
- EPSS
- 0.37% chance of exploitation in 30 days (percentile 23, 2026-11-26)
- CWE
- CWE-284
- Modified
Description
The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbitrary files or display incorrect information.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2022-3027. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2022-3027.md Affected: Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2022-3027 being exploited?
There is no confirmed exploitation: CVE-2022-3027 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.37% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2022-3027?
Medium: CVSS 5.7 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). A user has to take an action, such as opening a link, for the attack to work.
Which versions are affected by CVE-2022-3027?
Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor All.
Is there a fix for CVE-2022-3027?
No fixed version is listed in the sources yet. Vendor workaround: Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information. The following mitigations could assist in reducing the risk for exploitation of vulnerabilities: Disabling UART functionality at the CPU level Enforcing unique device authentication before granting access to the terminal / bootloader Where possible, enforcing secure boot. Tamper stickers on the device casing to indicate when a device has been opened
How do I check if I'm affected by CVE-2022-3027?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.