# CVE-2022-26388: Use of Hard-Coded Password Vulnerability in ELI Electrocardiograph Devices

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** MEDIUM 6.4
- **CVSS:** 6.4 (v3.1, CNA) `CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:L`
- **CVSS metrics:** Attack vector: Physical; Attack complexity: Low; Privileges required: None; User interaction: None; Scope: Unchanged; Confidentiality impact: High; Integrity impact: High; Availability impact: Low
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Not remotely exploitable (local or adjacent access needed)
- **EPSS:** 0.40% (percentile 26, 2026-11-26)
- **CWE:** CWE-259
- **Published:** 2025-02-07T17:06:30.134Z
- **Modified:** 2026-06-17T04:35:08.587Z

## Who should check

- Anyone running Welch Allyn ELI 380 Resting Electrocardiograph \<= 2.6.0
- Anyone running Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph \<= 2.3.1
- Anyone running Welch Allyn ELI 250c/BUR 250c Resting Electrocardiograph \<= 2.1.2
- Anyone running Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph \<= 2.2.0

## Description

A use of hard-coded password vulnerability may allow authentication abuse.This issue affects ELI 380 Resting Electrocardiograph: 

Versions 2.6.0 and prior; ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:

Versions 2.3.1 and prior; ELI 250c/BUR 250c Resting Electrocardiograph:

Versions 2.1.2 and prior; ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: 

Versions 2.2.0 and prior.

## Affected and fixed versions

- Welch Allyn ELI 380 Resting Electrocardiograph: \<= 2.6.0 (cvelist)
- Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: \<= 2.3.1 (cvelist)
- Welch Allyn ELI 250c/BUR 250c Resting Electrocardiograph: \<= 2.1.2 (cvelist)
- Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \<= 2.2.0 (cvelist)

## Mitigation

- Vendor workaround: Hillrom recommends the following workarounds to help reduce risk:
- 
- 
- 
-   \*  Apply proper network and physical security controls.
- 
-   \*  Ensure a unique encryption key is configured for ELI Link and Cardiograph.
- 
-   \*  Where possible, use a firewall to prevent communication on Port 21 
- FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet 
- service.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2022-26388. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2022-26388.md
Affected: Welch Allyn ELI 380 Resting Electrocardiograph <= 2.6.0; Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph <= 2.3.1; Welch Allyn ELI 250c/BUR 250c Resting Electrocardiograph <= 2.1.2; Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph <= 2.2.0
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Welch Allyn ELI 380 Resting Electrocardiograph, Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph, Welch Allyn ELI 250c/BUR 250c Resting Electrocardiograph, Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2022-26388 being exploited?

There is no confirmed exploitation: CVE-2022-26388 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.40% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2022-26388?

Medium: CVSS 6.4 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

### Which versions are affected by CVE-2022-26388?

Welch Allyn ELI 380 Resting Electrocardiograph \<= 2.6.0; Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph \<= 2.3.1; Welch Allyn ELI 250c/BUR 250c Resting Electrocardiograph \<= 2.1.2; Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph \<= 2.2.0.

### Is there a fix for CVE-2022-26388?

No fixed version is listed in the sources yet. Vendor workaround: Hillrom recommends the following workarounds to help reduce risk:      \*  Apply proper network and physical security controls.    \*  Ensure a unique encryption key is configured for ELI Link and Cardiograph.    \*  Where possible, use a firewall to prevent communication on Port 21  FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet  service.

### How do I check if I'm affected by CVE-2022-26388?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- other: <https://www.cisa.gov/news-events/ics-medical-advisories/icsma-22-167-01>
- other: <https://hillrom.com/en/responsible-disclosures/>

## Attribution

- CVE Program: CVE® is a registered trademark of The MITRE Corporation. CVE records © 1999–2026 The MITRE Corporation. <https://www.cve.org/Legal/TermsOfUse>
- NVD: This product uses data from the NVD API but is not endorsed or certified by the NVD. <https://nvd.nist.gov/vuln/detail/CVE-2022-26388>

HTML: <https://flawdb.com/CVE-2022-26388> · JSON: <https://flawdb.com/api/v1/entries/CVE-2022-26388>