CVE-2021-44464

MEDIUM 6.3EPSS 48%

Fresenius Kabi Agilia Connect Infusion System hard coded credentials

Mitigation

  • Upgrade Vigilant Software Suite (Mastermed Dashboard) to 2.0.1.3 or later.
  • Vendor workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
  • • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.
  • • Locate control system networks and remote devices behind firewalls and isolate them from the business network.
  • • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.

Who should check

  • Anyone running Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3

Vulnerability facts

Affected
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3
Fixed in
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Remote, login needed
CVSS
6.3 (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:LOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Network
  • Attack complexity: Low
  • Privileges required: Low
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: Low
  • Integrity impact: Low
  • Availability impact: Low
EPSS
0.68% chance of exploitation in 30 days (percentile 48, 2026-11-26)
CWE
CWE-798
Published
Modified

Description

Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software.

Affected and fixed versions

Package / productAffectedFixedSource
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard)>= unspecified, < 2.0.1.3—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2021-44464. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2021-44464.md
Affected: Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3
Fixed in: Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3

Steps:
1. Search this machine and project for Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard): dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2021-44464 being exploited?

There is no confirmed exploitation: CVE-2021-44464 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.68% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2021-44464?

Medium: CVSS 6.3 (v3.1, scored by CNA). Remote, login needed.

Which versions are affected by CVE-2021-44464?

Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3.

Is there a fix for CVE-2021-44464?

Yes. Fixed in Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3. Upgrade Vigilant Software Suite (Mastermed Dashboard) to 2.0.1.3 or later. Vendor workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet. • Locate control system networks and remote devices behind firewalls and isolate them from the business network. • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.

How do I check if I'm affected by CVE-2021-44464?