CVE-2021-44464
Fresenius Kabi Agilia Connect Infusion System hard coded credentials
Mitigation
- Upgrade Vigilant Software Suite (Mastermed Dashboard) to 2.0.1.3 or later.
- Vendor workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
- • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.
- • Locate control system networks and remote devices behind firewalls and isolate them from the business network.
- • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.
Who should check
- Anyone running Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3
Vulnerability facts
- Affected
- Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3
- Fixed in
- Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, login needed
- CVSS
- 6.3 (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:LOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Privileges required: Low
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: Low
- Integrity impact: Low
- Availability impact: Low
- EPSS
- 0.68% chance of exploitation in 30 days (percentile 48, 2026-11-26)
- CWE
- CWE-798
- Modified
Description
Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) | >= unspecified, < 2.0.1.3 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2021-44464. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2021-44464.md Affected: Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3 Fixed in: Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3 Steps: 1. Search this machine and project for Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard): dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2021-44464 being exploited?
There is no confirmed exploitation: CVE-2021-44464 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.68% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2021-44464?
Medium: CVSS 6.3 (v3.1, scored by CNA). Remote, login needed.
Which versions are affected by CVE-2021-44464?
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3.
Is there a fix for CVE-2021-44464?
Yes. Fixed in Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3. Upgrade Vigilant Software Suite (Mastermed Dashboard) to 2.0.1.3 or later. Vendor workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet. • Locate control system networks and remote devices behind firewalls and isolate them from the business network. • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.
How do I check if I'm affected by CVE-2021-44464?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.