Skip to content
FlawDB
Sign in
Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

CVE-2021-43935

HIGH 8.1EPSS 67%

ICSMA-21-343-01 Hillrom Welch Allyn Cardio Products

Mitigation

  • Vendor workaround: -Disable the SSO feature in the respective Modality Manager Configuration settings. Please refer to the instructions for use (IFU) and/or service manual for instructions on how to disable SSO.
  • -Apply proper network and physical security controls.
  • -Apply authentication for server access.

Who should check

  • Anyone running Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System >= 6.0.0, <= 6.3.1
  • Anyone running Hillrom Welch Allyn X-Scribe Cardiac Stress Testing System >= 5.01, <= 6.3.1
  • Anyone running Hillrom Welch Allyn Diagnostic Cardiology Suite 2.1.0
  • Anyone running Hillrom Welch Allyn Vision Express >= 6.1.0, <= 6.4.0
  • Anyone running Hillrom Welch Allyn H-Scribe Holter Analysis System >= 5.01, <= 6.4.0
  • Anyone running Hillrom Welch Allyn R-Scribe Resting ECG System >= 5.01, <= 7.0.0
  • Anyone running Hillrom Welch Allyn Connex Cardio >= 1.0.0, <= 1.1.1

Vulnerability facts

Affected
Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System >= 6.0.0, <= 6.3.1; Hillrom Welch Allyn X-Scribe Cardiac Stress Testing System >= 5.01, <= 6.3.1; Hillrom Welch Allyn Diagnostic Cardiology Suite 2.1.0; Hillrom Welch Allyn Vision Express >= 6.1.0, <= 6.4.0; Hillrom Welch Allyn H-Scribe Holter Analysis System >= 5.01, <= 6.4.0; Hillrom Welch Allyn R-Scribe Resting ECG System >= 5.01, <= 7.0.0; Hillrom Welch Allyn Connex Cardio >= 1.0.0, <= 1.1.1
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Remote, no login needed
CVSS
8.1 (v3.1, CNA)CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Network
  • Attack complexity: High
  • Privileges required: None
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: High
  • Integrity impact: High
  • Availability impact: High
EPSS
1.15% chance of exploitation in 30 days (percentile 67, 2026-11-26)
CWE
CWE-288
Published
Modified

Description

The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.

Affected and fixed versions

Package / productAffectedFixedSource
Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System>= 6.0.0, <= 6.3.1—cvelist
Hillrom Welch Allyn X-Scribe Cardiac Stress Testing System>= 5.01, <= 6.3.1—cvelist
Hillrom Welch Allyn Diagnostic Cardiology Suite2.1.0—cvelist
Hillrom Welch Allyn Vision Express>= 6.1.0, <= 6.4.0—cvelist
Hillrom Welch Allyn H-Scribe Holter Analysis System>= 5.01, <= 6.4.0—cvelist
Hillrom Welch Allyn R-Scribe Resting ECG System>= 5.01, <= 7.0.0—cvelist
Hillrom Welch Allyn Connex Cardio>= 1.0.0, <= 1.1.1—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2021-43935. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2021-43935.md
Affected: Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System >= 6.0.0, <= 6.3.1; Hillrom Welch Allyn X-Scribe Cardiac Stress Testing System >= 5.01, <= 6.3.1; Hillrom Welch Allyn Diagnostic Cardiology Suite 2.1.0; Hillrom Welch Allyn Vision Express >= 6.1.0, <= 6.4.0; Hillrom Welch Allyn H-Scribe Hol…
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System, Hillrom Welch Allyn X-Scribe Cardiac Stress Testing System, Hillrom Welch Allyn Diagnostic Cardiology Suite, Hillrom Welch Allyn Vision Express, Hillrom Welch Allyn H-Scribe Holter Analysis System, Hillrom Welch Allyn R-Scribe Resting ECG …: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2021-43935 being exploited?

There is no confirmed exploitation: CVE-2021-43935 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 1.15% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2021-43935?

High: CVSS 8.1 (v3.1, scored by CNA). Remote, no login needed. The attack needs conditions outside the attacker's control, which makes it harder.

Which versions are affected by CVE-2021-43935?

Hillrom Welch Allyn Q-Stress Cardiac Stress Testing System >= 6.0.0, <= 6.3.1; Hillrom Welch Allyn X-Scribe Cardiac Stress Testing System >= 5.01, <= 6.3.1; Hillrom Welch Allyn Diagnostic Cardiology Suite 2.1.0; Hillrom Welch Allyn Vision Express >= 6.1.0, <= 6.4.0; Hillrom Welch Allyn H-Scribe Holter Analysis System >= 5.01, <= 6.4.0; Hillrom Welch Allyn R-Scribe Resting ECG System >= 5.01, <= 7.0.0; Hillrom Welch Allyn Connex Cardio >= 1.0.0, <= 1.1.1.

Is there a fix for CVE-2021-43935?

No fixed version is listed in the sources yet. Vendor workaround: -Disable the SSO feature in the respective Modality Manager Configuration settings. Please refer to the instructions for use (IFU) and/or service manual for instructions on how to disable SSO. -Apply proper network and physical security controls. -Apply authentication for server access.

How do I check if I'm affected by CVE-2021-43935?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors