CVE-2021-38396

MEDIUM 6.5EPSS 4%

Missing Support Integrity Check for Boston Scientific Zoom Latitude

Mitigation

  • Vendor workaround: Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

Who should check

  • Anyone running Boston Scientific ZOOM LATITUDE Model 3120

Vulnerability facts

Affected
Boston Scientific ZOOM LATITUDE Model 3120
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
6.5 (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:LOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Physical
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: Required
  • Scope: Changed (impact reaches other components)
  • Confidentiality impact: Low
  • Integrity impact: High
  • Availability impact: Low
EPSS
0.17% chance of exploitation in 30 days (percentile 4, 2026-11-26)
CWE
CWE-353
Published
Modified

Description

The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.

Affected and fixed versions

Package / productAffectedFixedSource
Boston Scientific ZOOM LATITUDEModel 3120—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2021-38396. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2021-38396.md
Affected: Boston Scientific ZOOM LATITUDE Model 3120
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Boston Scientific ZOOM LATITUDE: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2021-38396 being exploited?

There is no confirmed exploitation: CVE-2021-38396 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.17% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2021-38396?

Medium: CVSS 6.5 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). A user has to take an action, such as opening a link, for the attack to work.

Which versions are affected by CVE-2021-38396?

Boston Scientific ZOOM LATITUDE Model 3120.

Is there a fix for CVE-2021-38396?

No fixed version is listed in the sources yet. Vendor workaround: Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120.

How do I check if I'm affected by CVE-2021-38396?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors