Skip to content
FlawDB
Sign in
Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

CVE-2021-23207

MEDIUM 6.5EPSS 8%

Fresenius Kabi Agilia Connect Infusion System plaintext storage of a password

Mitigation

  • Upgrade Vigilant Software Suite (Mastermed Dashboard) to 2.0.1.3 or later.
  • Upgrade Agilia Partner to 3.0 or later.
  • Vendor workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
  • • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.
  • • Locate control system networks and remote devices behind firewalls and isolate them from the business network.
  • • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.

Who should check

  • Anyone running Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3
  • Anyone running Fresenius Kabi Agilia Partner >= unspecified, < 3.0

Vulnerability facts

Affected
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3; Fresenius Kabi Agilia Partner >= unspecified, < 3.0
Fixed in
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3, Fresenius Kabi Agilia Partner 3.0
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
6.5 (v3.1, CNA)CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:C/C:H/I:N/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Local
  • Attack complexity: Low
  • Privileges required: Low
  • User interaction: None
  • Scope: Changed (impact reaches other components)
  • Confidentiality impact: High
  • Integrity impact: None
  • Availability impact: None
EPSS
0.23% chance of exploitation in 30 days (percentile 8, 2026-11-26)
CWE
CWE-256
Published
Modified

Description

An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users.

Affected and fixed versions

Package / productAffectedFixedSource
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard)>= unspecified, < 2.0.1.3—cvelist
Fresenius Kabi Agilia Partner>= unspecified, < 3.0—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2021-23207. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2021-23207.md
Affected: Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3; Fresenius Kabi Agilia Partner >= unspecified, < 3.0
Fixed in: Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3, Fresenius Kabi Agilia Partner 3.0

Steps:
1. Search this machine and project for Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard), Fresenius Kabi Agilia Partner: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2021-23207 being exploited?

There is no confirmed exploitation: CVE-2021-23207 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.23% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2021-23207?

Medium: CVSS 6.5 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

Which versions are affected by CVE-2021-23207?

Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) >= unspecified, < 2.0.1.3; Fresenius Kabi Agilia Partner >= unspecified, < 3.0.

Is there a fix for CVE-2021-23207?

Yes. Fixed in Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) 2.0.1.3, Fresenius Kabi Agilia Partner 3.0. Upgrade Vigilant Software Suite (Mastermed Dashboard) to 2.0.1.3 or later. Upgrade Agilia Partner to 3.0 or later. Vendor workaround: CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: • Minimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet. • Locate control system networks and remote devices behind firewalls and isolate them from the business network. • When remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.

How do I check if I'm affected by CVE-2021-23207?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.