In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Cente…

Mitigation

  • No fix listed yet: follow the vendor advisory.

Who should check

  • Anyone running GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 Monitors ApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3,CARESCAPE Central Station,v1.X,CARESCAPE Central Station,v2.X,B450,v2.X,B650,v1.X,B650,v2.X,B850,v1.X,B850,v2.X

Vulnerability facts

Affected
GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 Monitors ApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3,CARESCAPE Central Station,v1.X,CARESCAPE Central Station,v2.X,B450,v2.X,B650,v1.X,B650,v2.X,B850,v1.X,B850,v2.X
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Remote, no login needed
CVSS
8.6 (v3.1, NVD)CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Network
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: None
  • Scope: Changed (impact reaches other components)
  • Confidentiality impact: None
  • Integrity impact: High
  • Availability impact: None
EPSS
1.36% chance of exploitation in 30 days (percentile 72, 2026-11-26)
CWE
CWE-306
Published
Modified

Description

In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.

Affected and fixed versions

Package / productAffectedFixedSource
GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 MonitorsApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3,CARESCAPE Central Station,v1.X,CARESCAPE Central Station,v2.X,B450,v2.X,B650,v1.X,B650,v2.X,B850,v1.X,B850,v2.X—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2020-6964. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2020-6964.md
Affected: GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 Monitors ApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3…
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 Monitors: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2020-6964 being exploited?

There is no confirmed exploitation: CVE-2020-6964 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 1.36% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2020-6964?

High: CVSS 8.6 (v3.1, scored by NVD). Remote, no login needed.

Which versions are affected by CVE-2020-6964?

GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 Monitors ApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3,CARESCAPE Central Station,v1.X,CARESCAPE Central Station,v2.X,B450,v2.X,B650,v1.X,B650,v2.X,B850,v1.X,B850,v2.X.

Is there a fix for CVE-2020-6964?

No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.

How do I check if I'm affected by CVE-2020-6964?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors