# CVE-2020-25183: Medtronic MyCareLink Smart Improper Authentication

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** HIGH 8.0
- **CVSS:** 8.0 (v3.1, CNA) `CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H`
- **CVSS metrics:** Attack vector: Adjacent network; Attack complexity: Low; Privileges required: None; User interaction: Required; Scope: Unchanged; Confidentiality impact: High; Integrity impact: High; Availability impact: High
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Not remotely exploitable (local or adjacent access needed)
- **EPSS:** 0.80% (percentile 53, 2026-11-26)
- **CWE:** CWE-287
- **Published:** 2020-12-14T19:18:44.000Z
- **Modified:** 2026-06-17T03:06:32.203Z

## Who should check

- Anyone running Medtronic Smart Model 25000 Patient Reader All versions

## Description

Medtronic MyCareLink Smart 25000 contains 

an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

## Affected and fixed versions

- Medtronic Smart Model 25000 Patient Reader: All versions (cvelist)

## Mitigation

- Vendor workaround: In response to these vulnerabilities, Medtronic has applied additional controls for monitoring and responding to improper use of the MCL Smart Patient Reader:
- 
-   \*  Medtronic has implemented enhanced integrity validation (EIV) technology, which provides early detection and real-time mitigation of known vulnerability exploitation attempts.
-   \*  Medtronic has also implemented advanced detection system technology, which enables device-level logging and monitoring of all device activity and behavior.
- 
- 
- Medtronic recommends that users take additional defensive measures to minimize risk. Specifically, users should:
- 
-   \*  Maintain good physical control over home monitors.  \*  This includes only using home monitors in private environments such as a home, apartment, or otherwise physically controlled environment.
- 
- 
- 
-   \*  Use only home monitors obtained directly from your healthcare provider or a Medtronic representative.
-   \*  Patients should ensure that the operating system of their mobile phone is updated to the latest version of the available Android or Apple iOS operating system.
- 
- 
- Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2020-25183. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2020-25183.md
Affected: Medtronic Smart Model 25000 Patient Reader All versions
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Medtronic Smart Model 25000 Patient Reader: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2020-25183 being exploited?

There is no confirmed exploitation: CVE-2020-25183 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.80% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2020-25183?

High: CVSS 8.0 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). A user has to take an action, such as opening a link, for the attack to work.

### Which versions are affected by CVE-2020-25183?

Medtronic Smart Model 25000 Patient Reader All versions.

### Is there a fix for CVE-2020-25183?

No fixed version is listed in the sources yet. Vendor workaround: In response to these vulnerabilities, Medtronic has applied additional controls for monitoring and responding to improper use of the MCL Smart Patient Reader:    \*  Medtronic has implemented enhanced integrity validation (EIV) technology, which provides early detection and real-time mitigation of known vulnerability exploitation attempts.   \*  Medtronic has also implemented advanced detection system technology, which enables device-level logging and monitoring of all device activity and behavior.   Medtronic recommends that users take additional defensive measures to minimize risk. Specifically, users should:    \*  Maintain good physical control over home monitors.  \*  This includes only using home monitors in private environments such as a home, apartment, or otherwise physically controlled environment.      \*  Use only home monitors obtained directly from your healthcare provider or a Medtronic representative.   \*  Patients should ensure that the operating system of their mobile phone is updated to the latest version of the available Android or Apple iOS operating system.   Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative.

### How do I check if I'm affected by CVE-2020-25183?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- other: <https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-smart-security-vulnerability-patch.html>
- other: <https://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-345-01>
- other: <https://us-cert.cisa.gov/ics/advisories/icsma-20-345-01>