CVE-2020-25166
B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus
Mitigation
- Vendor workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms:
- Ensure the devices are not accessible directly from the Internet.
- Use a firewall and isolate the medical devices from the business network.
- Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
Who should check
- Anyone running B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81
- Anyone running B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81
- Anyone running B. Braun Melsungen AG Data module compactplus A10; A11
Vulnerability facts
- Affected
- B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Data module compactplus A10; A11
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, login needed
- CVSS
- 7.6 (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:LOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Privileges required: Low
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: Low
- Integrity impact: High
- Availability impact: Low
- EPSS
- 0.47% chance of exploitation in 30 days (percentile 33, 2026-11-26)
- CWE
- CWE-347
- Modified
Description
An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| B. Braun Melsungen AG SpaceCom | >= unspecified, <= U61; >= unspecified, <= L81 | — | cvelist |
| B. Braun Melsungen AG Battery pack with Wi-Fi | >= unspecified, <= U61; >= unspecified, <= L81 | — | cvelist |
| B. Braun Melsungen AG Data module compactplus | A10; A11 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2020-25166. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2020-25166.md Affected: B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Data module compactplus A10; A11 Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for B. Braun Melsungen AG SpaceCom, B. Braun Melsungen AG Battery pack with Wi-Fi, B. Braun Melsungen AG Data module compactplus: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2020-25166 being exploited?
There is no confirmed exploitation: CVE-2020-25166 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.47% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2020-25166?
High: CVSS 7.6 (v3.1, scored by CNA). Remote, login needed.
Which versions are affected by CVE-2020-25166?
B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Data module compactplus A10; A11.
Is there a fix for CVE-2020-25166?
No fixed version is listed in the sources yet. Vendor workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms: Ensure the devices are not accessible directly from the Internet. Use a firewall and isolate the medical devices from the business network. Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
How do I check if I'm affected by CVE-2020-25166?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.