Skip to content
FlawDB
Sign in
Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

CVE-2020-25154

MEDIUM 5.4EPSS 47%

B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

Mitigation

  • Vendor workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms:
  • Ensure the devices are not accessible directly from the Internet.
  • Use a firewall and isolate the medical devices from the business network.
  • Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

Who should check

  • Anyone running B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81
  • Anyone running B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81
  • Anyone running B. Braun Melsungen AG Data module compactplus A10; A11

Vulnerability facts

Affected
B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Data module compactplus A10; A11
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Remote, no login needed
CVSS
5.4 (v3.1, CNA)CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Network
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: Required
  • Scope: Unchanged
  • Confidentiality impact: Low
  • Integrity impact: Low
  • Availability impact: None
EPSS
0.66% chance of exploitation in 30 days (percentile 47, 2026-11-26)
CWE
CWE-601
Published
Modified

Description

An open redirect vulnerability in the administrative interface of the B. Braun Melsungen AG SpaceCom device Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to redirect users to malicious websites.

Affected and fixed versions

Package / productAffectedFixedSource
B. Braun Melsungen AG SpaceCom>= unspecified, <= U61; >= unspecified, <= L81—cvelist
B. Braun Melsungen AG Battery pack with Wi-Fi>= unspecified, <= U61; >= unspecified, <= L81—cvelist
B. Braun Melsungen AG Data module compactplusA10; A11—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2020-25154. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2020-25154.md
Affected: B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Data module compactplus A10; A11
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for B. Braun Melsungen AG SpaceCom, B. Braun Melsungen AG Battery pack with Wi-Fi, B. Braun Melsungen AG Data module compactplus: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2020-25154 being exploited?

There is no confirmed exploitation: CVE-2020-25154 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.66% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2020-25154?

Medium: CVSS 5.4 (v3.1, scored by CNA). Remote, no login needed. A user has to take an action, such as opening a link, for the attack to work.

Which versions are affected by CVE-2020-25154?

B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Data module compactplus A10; A11.

Is there a fix for CVE-2020-25154?

No fixed version is listed in the sources yet. Vendor workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms: Ensure the devices are not accessible directly from the Internet. Use a firewall and isolate the medical devices from the business network. Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

How do I check if I'm affected by CVE-2020-25154?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors