# CVE-2020-25152: B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** MEDIUM 6.5
- **CVSS:** 6.5 (v3.1, CNA) `CVSS:3.1/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N`
- **CVSS metrics:** Attack vector: Network; Attack complexity: Low; Privileges required: None; User interaction: Required; Scope: Unchanged; Confidentiality impact: High; Integrity impact: None; Availability impact: None
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Remote, no login needed
- **EPSS:** 1.28% (percentile 70, 2026-11-26)
- **CWE:** CWE-384
- **Published:** 2022-04-14T20:05:55.000Z
- **Modified:** 2026-06-17T03:06:27.680Z

## Who should check

- Anyone running B. Braun Melsungen AG SpaceCom \>= unspecified, \<= U61; \>= unspecified, \<= L81
- Anyone running B. Braun Melsungen AG Battery pack with Wi-Fi \>= unspecified, \<= U61; \>= unspecified, \<= L81
- Anyone running B. Braun Melsungen AG Data module compactplus A10; A11

## Description

A session fixation vulnerability in the B. Braun Melsungen AG SpaceCom administrative interface Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows remote attackers to hijack web sessions and escalate privileges.

## Affected and fixed versions

- B. Braun Melsungen AG SpaceCom: \>= unspecified, \<= U61; \>= unspecified, \<= L81 (cvelist)
- B. Braun Melsungen AG Battery pack with Wi-Fi: \>= unspecified, \<= U61; \>= unspecified, \<= L81 (cvelist)
- B. Braun Melsungen AG Data module compactplus: A10; A11 (cvelist)

## Mitigation

- Vendor workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms:
- 
-     Ensure the devices are not accessible directly from the Internet.
-     Use a firewall and isolate the medical devices from the business network. 
- 
- Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory.  https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2020-25152. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2020-25152.md
Affected: B. Braun Melsungen AG SpaceCom >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Battery pack with Wi-Fi >= unspecified, <= U61; >= unspecified, <= L81; B. Braun Melsungen AG Data module compactplus A10; A11
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for B. Braun Melsungen AG SpaceCom, B. Braun Melsungen AG Battery pack with Wi-Fi, B. Braun Melsungen AG Data module compactplus: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2020-25152 being exploited?

There is no confirmed exploitation: CVE-2020-25152 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 1.28% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2020-25152?

Medium: CVSS 6.5 (v3.1, scored by CNA). Remote, no login needed. A user has to take an action, such as opening a link, for the attack to work.

### Which versions are affected by CVE-2020-25152?

B. Braun Melsungen AG SpaceCom \>= unspecified, \<= U61; \>= unspecified, \<= L81; B. Braun Melsungen AG Battery pack with Wi-Fi \>= unspecified, \<= U61; \>= unspecified, \<= L81; B. Braun Melsungen AG Data module compactplus A10; A11.

### Is there a fix for CVE-2020-25152?

No fixed version is listed in the sources yet. Vendor workaround: As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms:      Ensure the devices are not accessible directly from the Internet.     Use a firewall and isolate the medical devices from the business network.   Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory.  https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

### How do I check if I'm affected by CVE-2020-25152?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- other: <https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02>
- other: <https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html>

## Attribution

- CVE Program: CVE® is a registered trademark of The MITRE Corporation. CVE records © 1999–2026 The MITRE Corporation. <https://www.cve.org/Legal/TermsOfUse>
- NVD: This product uses data from the NVD API but is not endorsed or certified by the NVD. <https://nvd.nist.gov/vuln/detail/CVE-2020-25152>

HTML: <https://flawdb.com/CVE-2020-25152> · JSON: <https://flawdb.com/api/v1/entries/CVE-2020-25152>