# CVE-2020-16218: Philips Patient Monitoring Devices Cross-site Scripting

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** LOW 3.5
- **CVSS:** 3.5 (v3.1, NVD) `CVSS:3.1/AV:A/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N`
- **CVSS metrics:** Attack vector: Adjacent network; Attack complexity: Low; Privileges required: Low; User interaction: None; Scope: Unchanged; Confidentiality impact: Low; Integrity impact: None; Availability impact: None
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Not remotely exploitable (local or adjacent access needed)
- **EPSS:** 0.65% (percentile 46, 2026-11-26)
- **CWE:** CWE-79
- **Published:** 2020-09-11T12:50:08.000Z
- **Modified:** 2026-06-17T02:57:52.740Z

## Who should check

- Anyone running Philips Patient Information Center iX (PICiX) B.02; C.02; C.03

## Description

In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the 
software does not neutralize or incorrectly neutralizes 
user-controllable input before it is placed in output that is then used 
as a webpage and served to other users. Successful exploitation could 
lead to unauthorized access to patient data via a read-only web 
application.

## Affected and fixed versions

- Philips Patient Information Center iX (PICiX): B.02; C.02; C.03 (cvelist)

## Mitigation

- Vendor workaround: As a mitigation to these vulnerabilities, Philips recommends the following:
- 
- 
- 
-   \*  The Philips patient monitoring network is required to be physically 
- or logically isolated from the hospital local area network (LAN). 
- Philips recommends using a firewall or routers that can implement access
-  control lists restricting access in and out of the patient monitoring 
- network for only necessary ports and IP addresses. Refer to the Philips 
- Patient Monitoring System Security for Clinical Networks guide for 
- additional information on  InCenter https://incenter.medical.philips.com/ .
- 
-   \*  By default, the simple certificate enrollment protocol (SCEP) 
- service is not running. When needed, the service is configured to run 
- based on the duration or the number of certificates to be assigned. One 
- certificate is default, but if a certificate is not issued, the service 
- will continue to run. Limit exposure by ensuring the SCEP service is not
-  running unless it is actively being used to enroll new devices.
- 
-   \*  When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits.
- 
-   \*  Implement physical security controls to prevent unauthorized login 
- attempts on the PIC iX application. Servers should be kept in controlled
-  locked data centers. Access to equipment at nurses’ stations should be 
- controlled and monitored.
- 
-   \*  Only grant remote access to PIC iX servers on a must-have basis.
- 
-   \*  Grant login privileges to the bedside monitor and PIC iX application
-  on a role-based, least-privilege basis, and only to trusted users.
- 
- 
- 
- 
- Users with questions regarding their specific Philips Patient 
- Information Center (PIC iX) and/or IntelliVue patient monitor 
- installations and new release eligibility should contact their local  Philips service support team, or regional service support https://www.usa.philips.com/healthcare/solutions/customer-service-solutions , or call 1-800-722-9377.
- 
- 
- Please see the  Philips product security website https://www.philips.com/productsecurity  for the Philips advisory and the latest security information for Philips products.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2020-16218. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2020-16218.md
Affected: Philips Patient Information Center iX (PICiX) B.02; C.02; C.03
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Philips Patient Information Center iX (PICiX): dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2020-16218 being exploited?

There is no confirmed exploitation: CVE-2020-16218 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.65% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2020-16218?

Low: CVSS 3.5 (v3.1, scored by NVD). Not remotely exploitable (local or adjacent access needed).

### Which versions are affected by CVE-2020-16218?

Philips Patient Information Center iX (PICiX) B.02; C.02; C.03.

### Is there a fix for CVE-2020-16218?

No fixed version is listed in the sources yet. Vendor workaround: As a mitigation to these vulnerabilities, Philips recommends the following:      \*  The Philips patient monitoring network is required to be physically  or logically isolated from the hospital local area network (LAN).  Philips recommends using a firewall or routers that can implement access  control lists restricting access in and out of the patient monitoring  network for only necessary ports and IP addresses. Refer to the Philips  Patient Monitoring System Security for Clinical Networks guide for  additional information on  InCenter https://incenter.medical.philips.com/ .    \*  By default, the simple certificate enrollment protocol (SCEP)  service is not running. When needed, the service is configured to run  based on the duration or the number of certificates to be assigned. One  certificate is default, but if a certificate is not issued, the service  will continue to run. Limit exposure by ensuring the SCEP service is not  running unless it is actively being used to enroll new devices.    \*  When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits.    \*  Implement physical security controls to prevent unauthorized login  attempts on the PIC iX application. Servers should be kept in controlled  locked data centers. Access to equipment at nurses’ stations should be  controlled and monitored.    \*  Only grant remote access to PIC iX servers on a must-have basis.    \*  Grant login privileges to the bedside monitor and PIC iX application  on a role-based, least-privilege basis, and only to trusted users.     Users with questions regarding their specific Philips Patient  Information Center (PIC iX) and/or IntelliVue patient monitor  installations and new release eligibility should contact their local  Philips service support team, or regional service support https://www.usa.philips.com/healthcare/solutions/customer-service-solutions , or call 1-800-722-9377.   Please see the  Philips product security website https://www.philips.com/productsecurity  for the Philips advisory and the latest security information for Philips products.

### How do I check if I'm affected by CVE-2020-16218?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- other: <https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01>
- other: <https://www.philips.com/productsecurity>