CVE-2020-16214
Philips Patient Monitoring Devices Improper Neutralization of Formula Elements in a CSV File
Mitigation
- Vendor workaround: As a mitigation to these vulnerabilities, Philips recommends the following:
- * The Philips patient monitoring network is required to be physically
- or logically isolated from the hospital local area network (LAN).
- Philips recommends using a firewall or routers that can implement access
- control lists restricting access in and out of the patient monitoring
- network for only necessary ports and IP addresses. Refer to the Philips
- Patient Monitoring System Security for Clinical Networks guide for
- additional information on InCenter https://incenter.medical.philips.com/ .
- * By default, the simple certificate enrollment protocol (SCEP)
- service is not running. When needed, the service is configured to run
- based on the duration or the number of certificates to be assigned. One
- certificate is default, but if a certificate is not issued, the service
- will continue to run. Limit exposure by ensuring the SCEP service is not
- running unless it is actively being used to enroll new devices.
- * When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits.
- * Implement physical security controls to prevent unauthorized login
- attempts on the PIC iX application. Servers should be kept in controlled
- locked data centers. Access to equipment at nurses’ stations should be
- controlled and monitored.
- * Only grant remote access to PIC iX servers on a must-have basis.
- * Grant login privileges to the bedside monitor and PIC iX application
- on a role-based, least-privilege basis, and only to trusted users.
- Users with questions regarding their specific Philips Patient
- Information Center (PIC iX) and/or IntelliVue patient monitor
- installations and new release eligibility should contact their local Philips service support team, or regional service support https://www.usa.philips.com/healthcare/solutions/customer-service-solutions , or call 1-800-722-9377.
- Please see the Philips product security website https://www.philips.com/productsecurity for the Philips advisory and the latest security information for Philips products.
Who should check
- Anyone running Philips Patient Information Center iX (PICiX) B.02; C.02; C.03
Vulnerability facts
- Affected
- Philips Patient Information Center iX (PICiX) B.02; C.02; C.03
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 5.0 (v3.1, NVD)CVSS:3.1/AV:L/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Local
- Attack complexity: Low
- Privileges required: None
- User interaction: Required
- Scope: Changed (impact reaches other components)
- Confidentiality impact: Low
- Integrity impact: Low
- Availability impact: None
- EPSS
- 0.59% chance of exploitation in 30 days (percentile 42, 2026-11-26)
- CWE
- CWE-1236
- Modified
Description
In Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, the software saves user-provided information into a comma-separated value (CSV) file, but it does not neutralize or incorrectly neutralizes special elements that could be interpreted as a command when the file is opened by spreadsheet software.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Philips Patient Information Center iX (PICiX) | B.02; C.02; C.03 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2020-16214. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2020-16214.md Affected: Philips Patient Information Center iX (PICiX) B.02; C.02; C.03 Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Philips Patient Information Center iX (PICiX): dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2020-16214 being exploited?
There is no confirmed exploitation: CVE-2020-16214 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.59% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2020-16214?
Medium: CVSS 5.0 (v3.1, scored by NVD). Not remotely exploitable (local or adjacent access needed). A user has to take an action, such as opening a link, for the attack to work.
Which versions are affected by CVE-2020-16214?
Philips Patient Information Center iX (PICiX) B.02; C.02; C.03.
Is there a fix for CVE-2020-16214?
No fixed version is listed in the sources yet. Vendor workaround: As a mitigation to these vulnerabilities, Philips recommends the following: * The Philips patient monitoring network is required to be physically or logically isolated from the hospital local area network (LAN). Philips recommends using a firewall or routers that can implement access control lists restricting access in and out of the patient monitoring network for only necessary ports and IP addresses. Refer to the Philips Patient Monitoring System Security for Clinical Networks guide for additional information on InCenter https://incenter.medical.philips.com/ . * By default, the simple certificate enrollment protocol (SCEP) service is not running. When needed, the service is configured to run based on the duration or the number of certificates to be assigned. One certificate is default, but if a certificate is not issued, the service will continue to run. Limit exposure by ensuring the SCEP service is not running unless it is actively being used to enroll new devices. * When enrolling new devices using SCEP, enter a unique challenge password of 8-12 unpredictable and randomized digits. * Implement physical security controls to prevent unauthorized login attempts on the PIC iX application. Servers should be kept in controlled locked data centers. Access to equipment at nurses’ stations should be controlled and monitored. * Only grant remote access to PIC iX servers on a must-have basis. * Grant login privileges to the bedside monitor and PIC iX application on a role-based, least-privilege basis, and only to trusted users. Users with questions regarding their specific Philips Patient Information Center (PIC iX) and/or IntelliVue patient monitor installations and new release eligibility should contact their local Philips service support team, or regional service support https://www.usa.philips.com/healthcare/solutions/customer-service-solutions , or call 1-800-722-9377. Please see the Philips product security website https://www.philips.com/productsecurity for the Philips advisory and the latest security information for Philips products.
How do I check if I'm affected by CVE-2020-16214?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.