CVE-2020-10627

HIGH 7.3EPSS 35%

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wi…

Mitigation

  • Vendor workaround: Insulet recommends patients using the affected products talk to their healthcare provider about the risks of continued use, including the possibility of changing to the latest model with increased cybersecurity protection.
  • Additionally, Insulet recommends all patients take the cybersecurity precautions indicated below.
  • Do not connect to or allow any third-party devices to be connected to or use any software not authorized by Insulet.
  • Maintain tight physical control of the pump and devices connected to the pump.
  • Be attentive to pump notifications, alarms, and alerts.
  • Immediately cancel any unintended boluses (a single dose of insulin administered all at once).
  • Monitor blood glucose levels closely and act as appropriate.
  • Get medical help immediately when experiencing symptoms of severe hypoglycemia or diabetic ketoacidosis or if you suspect insulin pump settings or insulin delivery has changed unexpectedly.
  • Insulet has released additional patient-focused information: https://www.myomnipod.com/security-bulletins
  • More information is available regarding Insulet’s product security and vulnerability management: https://www.myomnipod.com/product-security

Who should check

  • Anyone running Insulet Omnipod Insulin Management System 19191; 40160; ZXP425; ZXR425

Vulnerability facts

Affected
Insulet Omnipod Insulin Management System 19191; 40160; ZXP425; ZXR425
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
7.3 (v3.1, CNA)CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:LOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Local
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: Low
  • Integrity impact: High
  • Availability impact: Low
EPSS
0.49% chance of exploitation in 30 days (percentile 35, 2026-11-26)
CWE
CWE-284
Published
Modified

Description

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Affected and fixed versions

Package / productAffectedFixedSource
Insulet Omnipod Insulin Management System19191; 40160; ZXP425; ZXR425—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2020-10627. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2020-10627.md
Affected: Insulet Omnipod Insulin Management System 19191; 40160; ZXP425; ZXR425
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Insulet Omnipod Insulin Management System: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2020-10627 being exploited?

There is no confirmed exploitation: CVE-2020-10627 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.49% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2020-10627?

High: CVSS 7.3 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

Which versions are affected by CVE-2020-10627?

Insulet Omnipod Insulin Management System 19191; 40160; ZXP425; ZXR425.

Is there a fix for CVE-2020-10627?

No fixed version is listed in the sources yet. Vendor workaround: Insulet recommends patients using the affected products talk to their healthcare provider about the risks of continued use, including the possibility of changing to the latest model with increased cybersecurity protection. Additionally, Insulet recommends all patients take the cybersecurity precautions indicated below. Do not connect to or allow any third-party devices to be connected to or use any software not authorized by Insulet. Maintain tight physical control of the pump and devices connected to the pump. Be attentive to pump notifications, alarms, and alerts. Immediately cancel any unintended boluses (a single dose of insulin administered all at once). Monitor blood glucose levels closely and act as appropriate. Get medical help immediately when experiencing symptoms of severe hypoglycemia or diabetic ketoacidosis or if you suspect insulin pump settings or insulin delivery has changed unexpectedly. Insulet has released additional patient-focused information: https://www.myomnipod.com/security-bulletins More information is available regarding Insulet’s product security and vulnerability management: https://www.myomnipod.com/product-security

How do I check if I'm affected by CVE-2020-10627?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.