CVE-2019-6540
Medtronic Conexus Radio Frequency Telemetry Protocol Cleartext Transmission of Sensitive Information
Mitigation
- Vendor workaround: Medtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval.
- Medtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
- * Maintain good physical control over home monitors and programmers.
- * Use only home monitors, programmers, and implantable devices obtained directly from your healthcare provider or a Medtronic representative to ensure integrity of the system.
- * Do not connect unapproved devices to home monitors and programmers through USB ports or other physical connections.
- * Only use programmers to connect and interact with implanted devices in physically controlled hospital and clinical environments.
- * Only use home monitors in private environments such as a home, apartment, or otherwise physically controlled environment.
- * Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative.
- Medtronic has released additional patient focused information https://www.medtronic.com/security .
Who should check
- Anyone running Medtronic Conexus Radio Frequency Telemetry Protocol All versions
- Anyone running Medtronic MyCareLink Monitor 24950; 24952
- Anyone running Medtronic CareLink Monitor 2490C
- Anyone running Medtronic CareLink 2090 Programmer All versions
- Anyone running Medtronic Amplia CRT-D All versions
- Anyone running Medtronic Claria CRT-D All versions
- Anyone running Medtronic Compia CRT-D All versions
- Anyone running Medtronic Concerto CRT-D All versions
Vulnerability facts
- Affected
- Medtronic Conexus Radio Frequency Telemetry Protocol All versions; Medtronic MyCareLink Monitor 24950; 24952; Medtronic CareLink Monitor 2490C; Medtronic CareLink 2090 Programmer All versions; Medtronic Amplia CRT-D All versions; Medtronic Claria CRT-D All versions; Medtronic Compia CRT-D All versions; Medtronic Concerto CRT-D All versions; Medtronic Concerto II CRT-D All versions; Medtronic Consulta CRT-D All versions; Medtronic Evera ICD All versions; Medtronic Maximo II CRT-D All versions; Medtronic Maximo II ICD All versions; Medtronic Mirro ICD All versions; Medtronic Nayamed ND ICD All versions; Medtronic Primo ICD All versions; Medtronic Protecta ICD, Protecta CRT-D All versions; Medtronic Secura ICD All versions; Medtronic Virtuoso ICD All versions; Medtronic Virtuoso II ICD All versions; Medtronic Visia AF ICD All versions; Medtronic Viva CRT-D All versions; Medtronic Brava CRT-D All versions; Medtronic Mirro MRI ICD All versions
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 6.5 (v3.1, CNA)CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Adjacent network
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: High
- Integrity impact: None
- Availability impact: None
- EPSS
- 0.19% chance of exploitation in 30 days (percentile 5, 2026-11-26)
- CWE
- CWE-319
- Modified
Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Medtronic Conexus Radio Frequency Telemetry Protocol | All versions | — | cvelist |
| Medtronic MyCareLink Monitor | 24950; 24952 | — | cvelist |
| Medtronic CareLink Monitor | 2490C | — | cvelist |
| Medtronic CareLink 2090 Programmer | All versions | — | cvelist |
| Medtronic Amplia CRT-D | All versions | — | cvelist |
| Medtronic Claria CRT-D | All versions | — | cvelist |
| Medtronic Compia CRT-D | All versions | — | cvelist |
| Medtronic Concerto CRT-D | All versions | — | cvelist |
| Medtronic Concerto II CRT-D | All versions | — | cvelist |
| Medtronic Consulta CRT-D | All versions | — | cvelist |
| Medtronic Evera ICD | All versions | — | cvelist |
| Medtronic Maximo II CRT-D | All versions | — | cvelist |
| Medtronic Maximo II ICD | All versions | — | cvelist |
| Medtronic Mirro ICD | All versions | — | cvelist |
| Medtronic Nayamed ND ICD | All versions | — | cvelist |
| Medtronic Primo ICD | All versions | — | cvelist |
| Medtronic Protecta ICD, Protecta CRT-D | All versions | — | cvelist |
| Medtronic Secura ICD | All versions | — | cvelist |
| Medtronic Virtuoso ICD | All versions | — | cvelist |
| Medtronic Virtuoso II ICD | All versions | — | cvelist |
| Medtronic Visia AF ICD | All versions | — | cvelist |
| Medtronic Viva CRT-D | All versions | — | cvelist |
| Medtronic Brava CRT-D | All versions | — | cvelist |
| Medtronic Mirro MRI ICD | All versions | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2019-6540. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2019-6540.md Affected: Medtronic Conexus Radio Frequency Telemetry Protocol All versions; Medtronic MyCareLink Monitor 24950; 24952; Medtronic CareLink Monitor 2490C; Medtronic CareLink 2090 Programmer All versions; Medtronic Amplia CRT-D All versions; Medtronic Claria CRT-D All versions; Medtronic Compia CRT-D All versi… Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Medtronic Conexus Radio Frequency Telemetry Protocol, Medtronic MyCareLink Monitor, Medtronic CareLink Monitor, Medtronic CareLink 2090 Programmer, Medtronic Amplia CRT-D, Medtronic Claria CRT-D: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2019-6540 being exploited?
There is no confirmed exploitation: CVE-2019-6540 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.19% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2019-6540?
Medium: CVSS 6.5 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2019-6540?
Medtronic Conexus Radio Frequency Telemetry Protocol All versions; Medtronic MyCareLink Monitor 24950; 24952; Medtronic CareLink Monitor 2490C; Medtronic CareLink 2090 Programmer All versions; Medtronic Amplia CRT-D All versions; Medtronic Claria CRT-D All versions; Medtronic Compia CRT-D All versions; Medtronic Concerto CRT-D All versions; Medtronic Concerto II CRT-D All versions; Medtronic Consulta CRT-D All versions; Medtronic Evera ICD All versions; Medtronic Maximo II CRT-D All versions; Medtronic Maximo II ICD All versions; Medtronic Mirro ICD All versions; Medtronic Nayamed ND ICD All versions; Medtronic Primo ICD All versions; Medtronic Protecta ICD, Protecta CRT-D All versions; Medtronic Secura ICD All versions; Medtronic Virtuoso ICD All versions; Medtronic Virtuoso II ICD All versions; Medtronic Visia AF ICD All versions; Medtronic Viva CRT-D All versions; Medtronic Brava CRT-D All versions; Medtronic Mirro MRI ICD All versions.
Is there a fix for CVE-2019-6540?
No fixed version is listed in the sources yet. Vendor workaround: Medtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval. Medtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should: * Maintain good physical control over home monitors and programmers. * Use only home monitors, programmers, and implantable devices obtained directly from your healthcare provider or a Medtronic representative to ensure integrity of the system. * Do not connect unapproved devices to home monitors and programmers through USB ports or other physical connections. * Only use programmers to connect and interact with implanted devices in physically controlled hospital and clinical environments. * Only use home monitors in private environments such as a home, apartment, or otherwise physically controlled environment. * Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative. Medtronic has released additional patient focused information https://www.medtronic.com/security .
How do I check if I'm affected by CVE-2019-6540?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.