CVE-2019-25716
Dräger Infinity Delta/Kappa Patient Monitor DoS via Malformed Network Packet
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running Dräger Infinity Delta Infinity Delta
- Anyone running Dräger Infinity Delta XL Infinity Delta XL
- Anyone running Dräger Infinity Kappa Infinity Kappa
Vulnerability facts
- Affected
- Dräger Infinity Delta Infinity Delta; Dräger Infinity Delta XL Infinity Delta XL; Dräger Infinity Kappa Infinity Kappa
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 7.1 (v4.0, CNA)CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:NOpen in CVSS calculator
Show all 11 CVSS metrics
- Attack vector: Adjacent network
- Attack complexity: Low
- Attack requirements: None
- Privileges required: None
- User interaction: None
- Confidentiality impact: None
- Integrity impact: None
- Availability impact: High
- Confidentiality impact on other systems: None
- Integrity impact on other systems: None
- Availability impact on other systems: None
- EPSS
- 0.41% chance of exploitation in 30 days (percentile 27, 2026-11-26)
- CWE
- CWE-15
- Modified
Description
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
Affected and fixed versions
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2019-25716. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2019-25716.md Affected: Dräger Infinity Delta Infinity Delta; Dräger Infinity Delta XL Infinity Delta XL; Dräger Infinity Kappa Infinity Kappa Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Dräger Infinity Delta, Dräger Infinity Delta XL, Dräger Infinity Kappa: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
Vendor advisories
FAQ
Is CVE-2019-25716 being exploited?
There is no confirmed exploitation: CVE-2019-25716 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.41% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2019-25716?
High: CVSS 7.1 (v4.0, scored by CNA). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2019-25716?
Dräger Infinity Delta Infinity Delta; Dräger Infinity Delta XL Infinity Delta XL; Dräger Infinity Kappa Infinity Kappa.
Is there a fix for CVE-2019-25716?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2019-25716?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.