CVE-2019-18256
BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recovera…
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20
Vulnerability facts
- Affected
- BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 4.6 (v3.1, NVD)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Physical
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: High
- Integrity impact: None
- Availability impact: None
- EPSS
- 0.36% chance of exploitation in 30 days (percentile 22, 2026-11-26)
- CWE
- CWE-257
- Modified
Description
BIOTRONIK CardioMessenger II, The affected products use individual per-device credentials that are stored in a recoverable format. An attacker with physical access to the CardioMessenger can use these credentials for network authentication and decryption of local data in transit.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM | CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2019-18256. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2019-18256.md Affected: BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20 Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2019-18256 being exploited?
There is no confirmed exploitation: CVE-2019-18256 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.36% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2019-18256?
Medium: CVSS 4.6 (v3.1, scored by NVD). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2019-18256?
BIOTRONIK CardioMessenger II-S T-Line, CardioMessenger II-S GSM CardioMessenger II-S T-Line T4APP 2.20, CardioMessenger II-S GSM T4APP 2.20.
Is there a fix for CVE-2019-18256?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2019-18256?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.