CVE-2019-10964
Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control
Mitigation
- Vendor workaround: Medtronic recommends U.S. patients who are currently using the affected products talk to their healthcare provider about changing to a newer model insulin pump with increased cybersecurity protection. Patients outside the U.S. will receive a notification letter with instructions based on the country where they live.
- Medtronic recommends all patients take the cybersecurity precautions indicated below.
- CYBERSECURITY PRECAUTIONS RECOMMENDED FOR ALL PATIENTS:
- * Maintain tight physical control of the pump and devices connected to the pump
- * Do not share pump serial number
- * Be attentive to pump notifications, alarms, and alerts
- * Immediately cancel any unintended boluses (a single dose of insulin administered all at once)
- * Do not connect to any third-party devices or use any software not authorized by Medtronic
- * Disconnect CareLink USB devices from computers when not being used to download data from the pump
- * Monitor blood glucose levels closely and act as appropriate
- * Get medical help immediately when experiencing symptoms of severe hypoglycemia or diabetic ketoacidosis, or suspect an insulin pump settings, or insulin delivery have changed unexpectedly
- Medtronic has released additional patient-focused information, at the following location:
- https://www.medtronic.com/security
Who should check
- Anyone running Medtronic MiniMed 508 pump All versions
- Anyone running Medtronic MiniMed Paradigm 511 pump All versions
- Anyone running Medtronic MiniMed Paradigm 512/712 pumps All versions
- Anyone running Medtronic MiniMed Paradigm 712E pump All versions
- Anyone running Medtronic MiniMed Paradigm 515/715 pumps All versions
- Anyone running Medtronic MiniMed Paradigm 522/722 pumps All versions
- Anyone running Medtronic MiniMed Paradigm 522K/722K pumps All versions
- Anyone running Medtronic MiniMed Paradigm 523/723 pumps <= Software Versions 2.4A
Vulnerability facts
- Affected
- Medtronic MiniMed 508 pump All versions; Medtronic MiniMed Paradigm 511 pump All versions; Medtronic MiniMed Paradigm 512/712 pumps All versions; Medtronic MiniMed Paradigm 712E pump All versions; Medtronic MiniMed Paradigm 515/715 pumps All versions; Medtronic MiniMed Paradigm 522/722 pumps All versions; Medtronic MiniMed Paradigm 522K/722K pumps All versions; Medtronic MiniMed Paradigm 523/723 pumps <= Software Versions 2.4A; Medtronic MiniMed Paradigm 523K/723K pumps <= Software Versions 2.4A; Medtronic MiniMed Paradigm Veo 554/754 pumps <= Software Versions 2.6A; Medtronic MiniMed Paradigm Veo 554CM/754CM pumps <= Software Versions 2.7A
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 7.1 (v3.1, CNA)CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Adjacent network
- Attack complexity: High
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: Low
- Integrity impact: High
- Availability impact: High
- EPSS
- 1.23% chance of exploitation in 30 days (percentile 69, 2026-11-26)
- CWE
- CWE-284
- Modified
Description
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Medtronic MiniMed 508 pump | All versions | — | cvelist |
| Medtronic MiniMed Paradigm 511 pump | All versions | — | cvelist |
| Medtronic MiniMed Paradigm 512/712 pumps | All versions | — | cvelist |
| Medtronic MiniMed Paradigm 712E pump | All versions | — | cvelist |
| Medtronic MiniMed Paradigm 515/715 pumps | All versions | — | cvelist |
| Medtronic MiniMed Paradigm 522/722 pumps | All versions | — | cvelist |
| Medtronic MiniMed Paradigm 522K/722K pumps | All versions | — | cvelist |
| Medtronic MiniMed Paradigm 523/723 pumps | <= Software Versions 2.4A | — | cvelist |
| Medtronic MiniMed Paradigm 523K/723K pumps | <= Software Versions 2.4A | — | cvelist |
| Medtronic MiniMed Paradigm Veo 554/754 pumps | <= Software Versions 2.6A | — | cvelist |
| Medtronic MiniMed Paradigm Veo 554CM/754CM pumps | <= Software Versions 2.7A | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2019-10964. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2019-10964.md Affected: Medtronic MiniMed 508 pump All versions; Medtronic MiniMed Paradigm 511 pump All versions; Medtronic MiniMed Paradigm 512/712 pumps All versions; Medtronic MiniMed Paradigm 712E pump All versions; Medtronic MiniMed Paradigm 515/715 pumps All versions; Medtronic MiniMed Paradigm 522/722 pumps All ve… Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Medtronic MiniMed 508 pump, Medtronic MiniMed Paradigm 511 pump, Medtronic MiniMed Paradigm 512/712 pumps, Medtronic MiniMed Paradigm 712E pump, Medtronic MiniMed Paradigm 515/715 pumps, Medtronic MiniMed Paradigm 522/722 pumps: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2019-10964 being exploited?
There is no confirmed exploitation: CVE-2019-10964 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 1.23% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2019-10964?
High: CVSS 7.1 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). The attack needs conditions outside the attacker's control, which makes it harder.
Which versions are affected by CVE-2019-10964?
Medtronic MiniMed 508 pump All versions; Medtronic MiniMed Paradigm 511 pump All versions; Medtronic MiniMed Paradigm 512/712 pumps All versions; Medtronic MiniMed Paradigm 712E pump All versions; Medtronic MiniMed Paradigm 515/715 pumps All versions; Medtronic MiniMed Paradigm 522/722 pumps All versions; Medtronic MiniMed Paradigm 522K/722K pumps All versions; Medtronic MiniMed Paradigm 523/723 pumps <= Software Versions 2.4A; Medtronic MiniMed Paradigm 523K/723K pumps <= Software Versions 2.4A; Medtronic MiniMed Paradigm Veo 554/754 pumps <= Software Versions 2.6A; Medtronic MiniMed Paradigm Veo 554CM/754CM pumps <= Software Versions 2.7A.
Is there a fix for CVE-2019-10964?
No fixed version is listed in the sources yet. Vendor workaround: Medtronic recommends U.S. patients who are currently using the affected products talk to their healthcare provider about changing to a newer model insulin pump with increased cybersecurity protection. Patients outside the U.S. will receive a notification letter with instructions based on the country where they live. Medtronic recommends all patients take the cybersecurity precautions indicated below. CYBERSECURITY PRECAUTIONS RECOMMENDED FOR ALL PATIENTS: * Maintain tight physical control of the pump and devices connected to the pump * Do not share pump serial number * Be attentive to pump notifications, alarms, and alerts * Immediately cancel any unintended boluses (a single dose of insulin administered all at once) * Do not connect to any third-party devices or use any software not authorized by Medtronic * Disconnect CareLink USB devices from computers when not being used to download data from the pump * Monitor blood glucose levels closely and act as appropriate * Get medical help immediately when experiencing symptoms of severe hypoglycemia or diabetic ketoacidosis, or suspect an insulin pump settings, or insulin delivery have changed unexpectedly Medtronic has released additional patient-focused information, at the following location: https://www.medtronic.com/security
How do I check if I'm affected by CVE-2019-10964?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.