CVE-2018-5446
Medtronic 2090 Carelink Programmer Storing Passwords in a Recoverable Format
Mitigation
- Vendor workaround: Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual:
- * Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer.
- * Only connect the programmer to managed, secure networks.
- * Update the software on the programmer when Medtronic updates are available.
- * Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation.
- * Offline updates are available, contact your Medtronic representative for more information.
- Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism.
- Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800‑638‑1991.
- After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
- Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information.
- Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner.
- Medtronic has released a security bulletin for the 2090 CareLink Programmer https://www.medtronic.com/security .
Who should check
- Anyone running Medtronic 2090 CareLink Programmer All versions
- Anyone running Medtronic 29901 Encore Programmer All versions
Vulnerability facts
- Affected
- Medtronic 2090 CareLink Programmer All versions; Medtronic 29901 Encore Programmer All versions
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 4.9 (v3.1, CNA)CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Physical
- Attack complexity: High
- Privileges required: None
- User interaction: None
- Scope: Changed (impact reaches other components)
- Confidentiality impact: High
- Integrity impact: None
- Availability impact: None
- EPSS
- 0.39% chance of exploitation in 30 days (percentile 24, 2026-11-26)
- CWE
- CWE-257
- Modified
Description
Medtronic 2090 CareLink Programmer uses a per-product username and password that is stored in a recoverable format.
Affected and fixed versions
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2018-5446. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2018-5446.md Affected: Medtronic 2090 CareLink Programmer All versions; Medtronic 29901 Encore Programmer All versions Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Medtronic 2090 CareLink Programmer, Medtronic 29901 Encore Programmer: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2018-5446 being exploited?
There is no confirmed exploitation: CVE-2018-5446 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.39% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2018-5446?
Medium: CVSS 4.9 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). The attack needs conditions outside the attacker's control, which makes it harder.
Which versions are affected by CVE-2018-5446?
Medtronic 2090 CareLink Programmer All versions; Medtronic 29901 Encore Programmer All versions.
Is there a fix for CVE-2018-5446?
No fixed version is listed in the sources yet. Vendor workaround: Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual: * Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer. * Only connect the programmer to managed, secure networks. * Update the software on the programmer when Medtronic updates are available. * Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation. * Offline updates are available, contact your Medtronic representative for more information. Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism. Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800‑638‑1991. After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory. Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information. Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner. Medtronic has released a security bulletin for the 2090 CareLink Programmer https://www.medtronic.com/security .
How do I check if I'm affected by CVE-2018-5446?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.