CVE-2018-18984

MEDIUM 4.6EPSS 17%

Medtronic 9790, 2090 CareLink, and 29901 Encore Programmers Missing Encryption of Sensitive Data

Mitigation

  • Vendor workaround: The CareLink 9790 Programmer has been placed into end-of-life status and is no longer supported by Medtronic. Medtronic recommends users no longer use the 9790 for any purpose.
  • The CareLink 2090 and 29901 Encore programmers store PHI and PII as part of their normal operating procedure. Medtronic recommends that when devices are storing PHI/PII it should be retained on these programmers for the least amount of time necessary, and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy. Please contact a Medtronic representative for proper disposal and PHI/PII retention setting instructions.
  • All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer.
  • Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, hospitals and clinicians should:
  • * Maintain strict physical control of the programmer.
  • * Use only legitimately obtained programmers and not ones provided by any third party.
  • Proper disposal of these programmers and the associated electronic media storing data is critical for the continued protection of any PHI and PII residing on the programmer.
  • Medtronic has released a security bulletin related to this advisory that is available, with contact information, at the following location:
  • https://www.medtronic.com/security

Who should check

  • Anyone running Medtronic CareLink 9790 Programmer All versions
  • Anyone running Medtronic CareLink 2090 Programmer All versions
  • Anyone running Medtronic 29901 Encore Programmer All versions

Vulnerability facts

Affected
Medtronic CareLink 9790 Programmer All versions; Medtronic CareLink 2090 Programmer All versions; Medtronic 29901 Encore Programmer All versions
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
4.6 (v3.1, CNA)CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Physical
  • Attack complexity: Low
  • Privileges required: None
  • User interaction: None
  • Scope: Unchanged
  • Confidentiality impact: High
  • Integrity impact: None
  • Availability impact: None
EPSS
0.32% chance of exploitation in 30 days (percentile 17, 2026-11-26)
CWE
CWE-311
Published
Modified

Description

Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .

Affected and fixed versions

Package / productAffectedFixedSource
Medtronic CareLink 9790 ProgrammerAll versions—cvelist
Medtronic CareLink 2090 ProgrammerAll versions—cvelist
Medtronic 29901 Encore ProgrammerAll versions—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2018-18984. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2018-18984.md
Affected: Medtronic CareLink 9790 Programmer All versions; Medtronic CareLink 2090 Programmer All versions; Medtronic 29901 Encore Programmer All versions
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Medtronic CareLink 9790 Programmer, Medtronic CareLink 2090 Programmer, Medtronic 29901 Encore Programmer: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2018-18984 being exploited?

There is no confirmed exploitation: CVE-2018-18984 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.32% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2018-18984?

Medium: CVSS 4.6 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed).

Which versions are affected by CVE-2018-18984?

Medtronic CareLink 9790 Programmer All versions; Medtronic CareLink 2090 Programmer All versions; Medtronic 29901 Encore Programmer All versions.

Is there a fix for CVE-2018-18984?

No fixed version is listed in the sources yet. Vendor workaround: The CareLink 9790 Programmer has been placed into end-of-life status and is no longer supported by Medtronic. Medtronic recommends users no longer use the 9790 for any purpose. The CareLink 2090 and 29901 Encore programmers store PHI and PII as part of their normal operating procedure. Medtronic recommends that when devices are storing PHI/PII it should be retained on these programmers for the least amount of time necessary, and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy. Please contact a Medtronic representative for proper disposal and PHI/PII retention setting instructions. All affected programmers allow for the manual deletion of programmer-generated reports, which could contain PHI/PII. Medtronic recommends users delete these reports when no longer needed and prior to any disposition of the programmer. Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, hospitals and clinicians should: * Maintain strict physical control of the programmer. * Use only legitimately obtained programmers and not ones provided by any third party. Proper disposal of these programmers and the associated electronic media storing data is critical for the continued protection of any PHI and PII residing on the programmer. Medtronic has released a security bulletin related to this advisory that is available, with contact information, at the following location: https://www.medtronic.com/security

How do I check if I'm affected by CVE-2018-18984?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.