# CVE-2018-14786: Becton, Dickinson and Company (BD) Alaris Plus medical syringe pumps (models Alaris GS, Alaris GH, Alaris CC, and Alari…

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** CRITICAL 9.4
- **CVSS:** 9.4 (v3.0, NVD) `CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H`
- **CVSS metrics:** Attack vector: Network; Attack complexity: Low; Privileges required: None; User interaction: None; Scope: Unchanged; Confidentiality impact: Low; Integrity impact: High; Availability impact: High
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Remote, no login needed
- **EPSS:** 2.94% (percentile 88, 2026-11-26)
- **CWE:** CWE-287
- **Published:** 2018-08-23T19:00:00.000Z
- **Modified:** 2026-06-17T01:41:38.287Z

## Who should check

- Anyone running Becton, Dickinson and Company Alaris GS, Alaris GH, Alaris CC, and Alaris TIVA Version 2.3.6 and prior

## Description

Becton, Dickinson and Company (BD) Alaris Plus medical syringe pumps (models Alaris GS, Alaris GH, Alaris CC, and Alaris TIVA) versions 2.3.6 and prior are affected by an improper authentication vulnerability where the software does not perform authentication for functionality that requires a provable user identity, where it may allow a remote attacker to gain unauthorized access to various Alaris Syringe pumps and impact the intended operation of the pump when it is connected to a terminal server via the serial port.

## Affected and fixed versions

- Becton, Dickinson and Company Alaris GS, Alaris GH, Alaris CC, and Alaris TIVA: Version 2.3.6 and prior (cvelist)

## Mitigation

- No fix listed yet: follow the vendor advisory.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2018-14786. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2018-14786.md
Affected: Becton, Dickinson and Company Alaris GS, Alaris GH, Alaris CC, and Alaris TIVA Version 2.3.6 and prior
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Becton, Dickinson and Company Alaris GS, Alaris GH, Alaris CC, and Alaris TIVA: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2018-14786 being exploited?

There is no confirmed exploitation: CVE-2018-14786 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 2.94% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2018-14786?

Critical: CVSS 9.4 (v3.0, scored by NVD). Remote, no login needed.

### Which versions are affected by CVE-2018-14786?

Becton, Dickinson and Company Alaris GS, Alaris GH, Alaris CC, and Alaris TIVA Version 2.3.6 and prior.

### Is there a fix for CVE-2018-14786?

No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.

### How do I check if I'm affected by CVE-2018-14786?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- vendor: <http://www.bd.com/en-us/support/product-security-and-privacy/product-security-bulletin-for-various-alaris-plus-syringe-pumps-sold-and-in-use-outside-the-united-states>
- other: <http://www.securityfocus.com/bid/105147>
- other: <https://ics-cert.us-cert.gov/advisories/ICSMA-18-235-01>