CVE-2018-14781
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
Mitigation
- Vendor workaround: The remote option is turned off in the pump by default.
- Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic.
- Medtronic has released additional patient focused information https://www.medtronic.com/security .
- Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic.
Who should check
- Anyone running Medtronic MMT- 508 - MiniMed pump All versions
- Anyone running Medtronic MMT – 511 pump Paradigm All versions
- Anyone running Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions
- Anyone running Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions
- Anyone running Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions
- Anyone running Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME All versions
- Anyone running Medtronic MMT – 523 / MMT – 723 Paradigm Revel All versions
- Anyone running Medtronic MMT – 523(K) / MMT – 723(K) Paradigm All versions
Vulnerability facts
- Affected
- Medtronic MMT- 508 - MiniMed pump All versions; Medtronic MMT – 511 pump Paradigm All versions; Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions; Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions; Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions; Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME All versions; Medtronic MMT – 523 / MMT – 723 Paradigm Revel All versions; Medtronic MMT – 523(K) / MMT – 723(K) Paradigm All versions; Medtronic MMT – 554 / MMT – 754 MiniMed Veo All versions; Medtronic MMT – 551 / MMT – 751 MiniMed 530G All versions
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 5.3 (v3.1, CNA)CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Adjacent network
- Attack complexity: High
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: None
- Integrity impact: High
- Availability impact: None
- EPSS
- 0.71% chance of exploitation in 30 days (percentile 49, 2026-11-26)
- CWE
- CWE-294
- Modified
Description
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Medtronic MMT- 508 - MiniMed pump | All versions | — | cvelist |
| Medtronic MMT – 511 pump Paradigm | All versions | — | cvelist |
| Medtronic MMT – 512 / MMT – 712 Paradigm x12 | All versions | — | cvelist |
| Medtronic MMT – 515 / MMT – 715 Paradigm x15 | All versions | — | cvelist |
| Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME | All versions | — | cvelist |
| Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME | All versions | — | cvelist |
| Medtronic MMT – 523 / MMT – 723 Paradigm Revel | All versions | — | cvelist |
| Medtronic MMT – 523(K) / MMT – 723(K) Paradigm | All versions | — | cvelist |
| Medtronic MMT – 554 / MMT – 754 MiniMed Veo | All versions | — | cvelist |
| Medtronic MMT – 551 / MMT – 751 MiniMed 530G | All versions | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2018-14781. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2018-14781.md Affected: Medtronic MMT- 508 - MiniMed pump All versions; Medtronic MMT – 511 pump Paradigm All versions; Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions; Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions; Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions; Medtronic MMT – 522(… Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Medtronic MMT- 508 - MiniMed pump, Medtronic MMT – 511 pump Paradigm, Medtronic MMT – 512 / MMT – 712 Paradigm x12, Medtronic MMT – 515 / MMT – 715 Paradigm x15, Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME, Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2018-14781 being exploited?
There is no confirmed exploitation: CVE-2018-14781 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.71% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2018-14781?
Medium: CVSS 5.3 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). The attack needs conditions outside the attacker's control, which makes it harder.
Which versions are affected by CVE-2018-14781?
Medtronic MMT- 508 - MiniMed pump All versions; Medtronic MMT – 511 pump Paradigm All versions; Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions; Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions; Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions; Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME All versions; Medtronic MMT – 523 / MMT – 723 Paradigm Revel All versions; Medtronic MMT – 523(K) / MMT – 723(K) Paradigm All versions; Medtronic MMT – 554 / MMT – 754 MiniMed Veo All versions; Medtronic MMT – 551 / MMT – 751 MiniMed 530G All versions.
Is there a fix for CVE-2018-14781?
No fixed version is listed in the sources yet. Vendor workaround: The remote option is turned off in the pump by default. Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic. Medtronic has released additional patient focused information https://www.medtronic.com/security . Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic.
How do I check if I'm affected by CVE-2018-14781?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.