CVE-2018-10634

MEDIUM 4.8EPSS 33%

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Cleartext Transmission of Sensitive Information

Mitigation

  • Vendor workaround: The remote option is turned off in the pump by default.  
  • Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic.
  • Medtronic has released additional patient focused information https://www.medtronic.com/security .
  • Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic.

Who should check

  • Anyone running Medtronic MMT- 508 - MiniMed pump All versions
  • Anyone running Medtronic MMT – 511 pump Paradigm All versions
  • Anyone running Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions
  • Anyone running Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions
  • Anyone running Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions
  • Anyone running Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME All versions
  • Anyone running Medtronic MMT – 523 / MMT – 723 Paradigm Revel All versions
  • Anyone running Medtronic MMT – 523(K) / MMT – 723(K) Paradigm All versions

Vulnerability facts

Affected
Medtronic MMT- 508 - MiniMed pump All versions; Medtronic MMT – 511 pump Paradigm All versions; Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions; Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions; Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions; Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME All versions; Medtronic MMT – 523 / MMT – 723 Paradigm Revel All versions; Medtronic MMT – 523(K) / MMT – 723(K) Paradigm All versions; Medtronic MMT – 554 / MMT – 754 MiniMed Veo All versions; Medtronic MMT – 551 / MMT – 751 MiniMed 530G All versions
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
4.8 (v3.1, CNA)CVSS:3.1/AV:A/AC:H/PR:N/UI:R/S:U/C:H/I:N/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Adjacent network
  • Attack complexity: High
  • Privileges required: None
  • User interaction: Required
  • Scope: Unchanged
  • Confidentiality impact: High
  • Integrity impact: None
  • Availability impact: None
EPSS
0.47% chance of exploitation in 30 days (percentile 33, 2026-11-26)
CWE
CWE-319
Published
Modified

Description

Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.

Affected and fixed versions

Package / productAffectedFixedSource
Medtronic MMT- 508 - MiniMed pumpAll versions—cvelist
Medtronic MMT – 511 pump ParadigmAll versions—cvelist
Medtronic MMT – 512 / MMT – 712 Paradigm x12All versions—cvelist
Medtronic MMT – 515 / MMT – 715 Paradigm x15All versions—cvelist
Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIMEAll versions—cvelist
Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIMEAll versions—cvelist
Medtronic MMT – 523 / MMT – 723 Paradigm RevelAll versions—cvelist
Medtronic MMT – 523(K) / MMT – 723(K) ParadigmAll versions—cvelist
Medtronic MMT – 554 / MMT – 754 MiniMed VeoAll versions—cvelist
Medtronic MMT – 551 / MMT – 751 MiniMed 530GAll versions—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2018-10634. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2018-10634.md
Affected: Medtronic MMT- 508 - MiniMed pump All versions; Medtronic MMT – 511 pump Paradigm All versions; Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions; Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions; Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions; Medtronic MMT – 522(…
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Medtronic MMT- 508 - MiniMed pump, Medtronic MMT – 511 pump Paradigm, Medtronic MMT – 512 / MMT – 712 Paradigm x12, Medtronic MMT – 515 / MMT – 715 Paradigm x15, Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME, Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2018-10634 being exploited?

There is no confirmed exploitation: CVE-2018-10634 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.47% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2018-10634?

Medium: CVSS 4.8 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). A user has to take an action, such as opening a link, for the attack to work. The attack needs conditions outside the attacker's control, which makes it harder.

Which versions are affected by CVE-2018-10634?

Medtronic MMT- 508 - MiniMed pump All versions; Medtronic MMT – 511 pump Paradigm All versions; Medtronic MMT – 512 / MMT – 712 Paradigm x12 All versions; Medtronic MMT – 515 / MMT – 715 Paradigm x15 All versions; Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME All versions; Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME All versions; Medtronic MMT – 523 / MMT – 723 Paradigm Revel All versions; Medtronic MMT – 523(K) / MMT – 723(K) Paradigm All versions; Medtronic MMT – 554 / MMT – 754 MiniMed Veo All versions; Medtronic MMT – 551 / MMT – 751 MiniMed 530G All versions.

Is there a fix for CVE-2018-10634?

No fixed version is listed in the sources yet. Vendor workaround: The remote option is turned off in the pump by default.   Medtronic is directing all users to stop using their remote controllers, disable the remote option on their insulin pump, and to return the remote controllers to Medtronic. Medtronic has released additional patient focused information https://www.medtronic.com/security . Additionally, Medtronic will be sending a letter to patients who may still be actively using the remotes in order to inform patients about these security risks, and request patients stop using the remote and return them to Medtronic.

How do I check if I'm affected by CVE-2018-10634?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

Sponsors