MEDIUM 6.3EPSS 27%

Medtronic N'Vision Clinician Programmer Protection Mechanism Failure

Mitigation

  • Vendor workaround: Medtronic has not developed a product update to address the vulnerabilities, but is reinforcing security reminders within this advisory to help reduce the risk associated with the vulnerabilities.
  • The 8870 Therapy Application card stores PHI and PII as part of its normal operating procedure and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy.
  • Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, hospitals and clinicians should:
  • * Maintain strict physical control of the 8870 application card.
  • * Use only legitimately obtained 8870 cards and not cards provided by any third party as firmware and system updates are provided directly by Medtronic using new 8870 application cards.
  • * 8840 Programmers and 8870 Therapy Application compact flash cards are the property of Medtronic and should be returned to Medtronic when no longer in use. If that is not an option, you should securely dispose of them.
  • Medtronic has released additional patient focused information, at the following location:
  • https://www.medtronic.com/security

Who should check

  • Anyone running Medtronic 8840 N’Vision Clinician Programmer All versions
  • Anyone running Medtronic 8870 N’Vision removable Application Card All versions

Vulnerability facts

Affected
Medtronic 8840 N’Vision Clinician Programmer All versions; Medtronic 8870 N’Vision removable Application Card All versions
Fixed in
No fix listed yet
Exploited in the wild
Not in CISA KEV
Public exploit / PoC
None known
Remote exploitation
Not remotely exploitable (local or adjacent access needed)
CVSS
6.3 (v3.1, CNA)CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
  • Attack vector: Physical
  • Attack complexity: High
  • Privileges required: None
  • User interaction: Required
  • Scope: Unchanged
  • Confidentiality impact: High
  • Integrity impact: High
  • Availability impact: High
EPSS
0.41% chance of exploitation in 30 days (percentile 27, 2026-11-26)
CWE
CWE-693
Published
Modified

Description

The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.

Affected and fixed versions

Package / productAffectedFixedSource
Medtronic 8840 N’Vision Clinician ProgrammerAll versions—cvelist
Medtronic 8870 N’Vision removable Application CardAll versions—cvelist

Check your systems with an agent

Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.

Check whether anything I run is affected by CVE-2018-10631. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2018-10631.md
Affected: Medtronic 8840 N’Vision Clinician Programmer All versions; Medtronic 8870 N’Vision removable Application Card All versions
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Medtronic 8840 N’Vision Clinician Programmer, Medtronic 8870 N’Vision removable Application Card: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.

References

FAQ

Is CVE-2018-10631 being exploited?

There is no confirmed exploitation: CVE-2018-10631 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.41% chance of exploitation in the next 30 days (as of 2026-11-26).

How serious is CVE-2018-10631?

Medium: CVSS 6.3 (v3.1, scored by CNA). Not remotely exploitable (local or adjacent access needed). A user has to take an action, such as opening a link, for the attack to work. The attack needs conditions outside the attacker's control, which makes it harder.

Which versions are affected by CVE-2018-10631?

Medtronic 8840 N’Vision Clinician Programmer All versions; Medtronic 8870 N’Vision removable Application Card All versions.

Is there a fix for CVE-2018-10631?

No fixed version is listed in the sources yet. Vendor workaround: Medtronic has not developed a product update to address the vulnerabilities, but is reinforcing security reminders within this advisory to help reduce the risk associated with the vulnerabilities. The 8870 Therapy Application card stores PHI and PII as part of its normal operating procedure and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy. Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, hospitals and clinicians should: * Maintain strict physical control of the 8870 application card. * Use only legitimately obtained 8870 cards and not cards provided by any third party as firmware and system updates are provided directly by Medtronic using new 8870 application cards. * 8840 Programmers and 8870 Therapy Application compact flash cards are the property of Medtronic and should be returned to Medtronic when no longer in use. If that is not an option, you should securely dispose of them. Medtronic has released additional patient focused information, at the following location: https://www.medtronic.com/security

How do I check if I'm affected by CVE-2018-10631?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.