CVE-2018-10601
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Moni…
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3
Vulnerability facts
- Affected
- ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 8.2 (v3.1, NVD)CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Adjacent network
- Attack complexity: High
- Privileges required: None
- User interaction: None
- Scope: Changed (impact reaches other components)
- Confidentiality impact: High
- Integrity impact: Low
- Availability impact: High
- EPSS
- 0.37% chance of exploitation in 30 days (percentile 23, 2026-11-26)
- CWE
- CWE-121
- Modified
Description
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors | The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2018-10601. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2018-10601.md Affected: ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 f… Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2018-10601 being exploited?
There is no confirmed exploitation: CVE-2018-10601 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.37% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2018-10601?
High: CVSS 8.2 (v3.1, scored by NVD). Not remotely exploitable (local or adjacent access needed). The attack needs conditions outside the attacker's control, which makes it harder.
Which versions are affected by CVE-2018-10601?
ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors The following IntelliVue Patient Monitors versions are affected: IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only). The following Avalon Fetal/Maternal Monitors versions are affected: Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3.
Is there a fix for CVE-2018-10601?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2018-10601?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.