# CVE-2018-10596: Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** HIGH 7.1
- **CVSS:** 7.1 (v3.1, CNA) `CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H`
- **CVSS metrics:** Attack vector: Network; Attack complexity: High; Privileges required: Low; User interaction: Required; Scope: Unchanged; Confidentiality impact: High; Integrity impact: High; Availability impact: High
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Remote, login needed
- **EPSS:** 1.33% (percentile 71, 2026-11-26)
- **CWE:** CWE-923
- **Published:** 2018-07-02T18:00:00.000Z
- **Modified:** 2026-06-17T01:34:14.603Z

## Who should check

- Anyone running Medtronic 2090 CareLink Programmer All versions
- Anyone running Medtronic 29901 Encore Programmer All versions

## Description

Medtronic 2090 CareLink Programmer 

uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

## Affected and fixed versions

- Medtronic 2090 CareLink Programmer: All versions (cvelist)
- Medtronic 29901 Encore Programmer: All versions (cvelist)

## Mitigation

- Vendor workaround: Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual:
- 
-   \*  Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer.
-   \*  Only connect the programmer to managed, secure networks.
-   \*  Update the software on the programmer when Medtronic updates are available.
-   \*  Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation. 
-   \*  Offline updates are available, contact your Medtronic representative for more information.
- 
- 
- Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism.
- 
- Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800‑638‑1991.
- 
- 
- After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.
- 
- Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information.
- 
- Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner.
- 
- Medtronic has released a security bulletin for the  2090 CareLink Programmer https://www.medtronic.com/security .

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2018-10596. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2018-10596.md
Affected: Medtronic 2090 CareLink Programmer All versions; Medtronic 29901 Encore Programmer All versions
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Medtronic 2090 CareLink Programmer, Medtronic 29901 Encore Programmer: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2018-10596 being exploited?

There is no confirmed exploitation: CVE-2018-10596 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 1.33% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2018-10596?

High: CVSS 7.1 (v3.1, scored by CNA). Remote, login needed. A user has to take an action, such as opening a link, for the attack to work. The attack needs conditions outside the attacker's control, which makes it harder.

### Which versions are affected by CVE-2018-10596?

Medtronic 2090 CareLink Programmer All versions; Medtronic 29901 Encore Programmer All versions.

### Is there a fix for CVE-2018-10596?

No fixed version is listed in the sources yet. Vendor workaround: Medtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual:    \*  Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer.   \*  Only connect the programmer to managed, secure networks.   \*  Update the software on the programmer when Medtronic updates are available.   \*  Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation.    \*  Offline updates are available, contact your Medtronic representative for more information.   Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism.  Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800‑638‑1991.   After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory.  Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information.  Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner.  Medtronic has released a security bulletin for the  2090 CareLink Programmer https://www.medtronic.com/security .

### How do I check if I'm affected by CVE-2018-10596?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- other: <https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html>
- other: <https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01>
- other: <https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01>

## Attribution

- CVE Program: CVE® is a registered trademark of The MITRE Corporation. CVE records © 1999–2026 The MITRE Corporation. <https://www.cve.org/Legal/TermsOfUse>
- NVD: This product uses data from the NVD API but is not endorsed or certified by the NVD. <https://nvd.nist.gov/vuln/detail/CVE-2018-10596>

HTML: <https://flawdb.com/CVE-2018-10596> · JSON: <https://flawdb.com/api/v1/entries/CVE-2018-10596>