CVE-2017-14018
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions…
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running Ethicon Endo-Surgery Generator G11 Ethicon Endo-Surgery Generator G11
Vulnerability facts
- Affected
- Ethicon Endo-Surgery Generator G11 Ethicon Endo-Surgery Generator G11
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 4.8 (v3.0, NVD)CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:LOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Physical
- Attack complexity: High
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: None
- Integrity impact: High
- Availability impact: Low
- EPSS
- 0.36% chance of exploitation in 30 days (percentile 22, 2026-11-26)
- CWE
- CWE-287
- Modified
Description
An improper authentication issue was discovered in Johnson & Johnson Ethicon Endo-Surgery Generator Gen11, all versions released before November 29, 2017. The security authentication mechanism used between the Ethicon Endo-Surgery Generator Gen11 and single-patient use products can be bypassed, allowing for unauthorized devices to be connected to the generator, which could result in a loss of integrity or availability.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Ethicon Endo-Surgery Generator G11 | Ethicon Endo-Surgery Generator G11 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2017-14018. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2017-14018.md Affected: Ethicon Endo-Surgery Generator G11 Ethicon Endo-Surgery Generator G11 Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Ethicon Endo-Surgery Generator G11: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2017-14018 being exploited?
There is no confirmed exploitation: CVE-2017-14018 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.36% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2017-14018?
Medium: CVSS 4.8 (v3.0, scored by NVD). Not remotely exploitable (local or adjacent access needed). The attack needs conditions outside the attacker's control, which makes it harder.
Which versions are affected by CVE-2017-14018?
Ethicon Endo-Surgery Generator G11 Ethicon Endo-Surgery Generator G11.
Is there a fix for CVE-2017-14018?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2017-14018?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.