CVE-2015-1012
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira,…
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running Hospira LifeCare PCA Infusion System <= 5.0
Vulnerability facts
- Affected
- Hospira LifeCare PCA Infusion System <= 5.0
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Remote, no login needed
- CVSS
- 7.5 (v3.0, NVD)CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:NOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Network
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: High
- Integrity impact: None
- Availability impact: None
- EPSS
- 0.77% chance of exploitation in 30 days (percentile 52, 2026-11-26)
- CWE
- CWE-312
- Modified
Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Hospira LifeCare PCA Infusion System | <= 5.0 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2015-1012. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2015-1012.md Affected: Hospira LifeCare PCA Infusion System <= 5.0 Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Hospira LifeCare PCA Infusion System: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2015-1012 being exploited?
There is no confirmed exploitation: CVE-2015-1012 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.77% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2015-1012?
High: CVSS 7.5 (v3.0, scored by NVD). Remote, no login needed.
Which versions are affected by CVE-2015-1012?
Hospira LifeCare PCA Infusion System <= 5.0.
Is there a fix for CVE-2015-1012?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2015-1012?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.