# CVE-2014-5432: Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is re…

> Archived entry. Full-text search covers entries since 2026-11-10; exact IDs work for all years.

- **Severity:** CRITICAL 9.8
- **CVSS:** 9.8 (v3.0, NVD) `CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H`
- **CVSS metrics:** Attack vector: Network; Attack complexity: Low; Privileges required: None; User interaction: None; Scope: Unchanged; Confidentiality impact: High; Integrity impact: High; Availability impact: High
- **Exploited in the wild:** Not in CISA KEV
- **Public exploit / PoC:** None known
- **Remote exploitation:** Remote, no login needed
- **EPSS:** 2.57% (percentile 86, 2026-11-26)
- **CWE:** CWE-592
- **Published:** 2019-03-26T15:17:16.000Z
- **Modified:** 2026-06-17T00:11:32.940Z

## Who should check

- Anyone running Baxter SIGMA Spectrum Infusion System 6.05 (model 35700BAX) with wireless battery module (WBM) version 16.

## Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

## Affected and fixed versions

- Baxter SIGMA Spectrum Infusion System: 6.05 (model 35700BAX) with wireless battery module (WBM) version 16. (cvelist)

## Mitigation

- No fix listed yet: follow the vendor advisory.

## Check your systems with an agent

Paste this into a coding agent:

```text
Check whether anything I run is affected by CVE-2014-5432. Work read-only: do not install, upgrade, delete or change anything.

Details: https://flawdb.com/CVE-2014-5432.md
Affected: Baxter SIGMA Spectrum Infusion System 6.05 (model 35700BAX) with wireless battery module (WBM) version 16.
Fixed in: no fixed version listed yet

Steps:
1. Search this machine and project for Baxter SIGMA Spectrum Infusion System: dependency manifests and lockfiles, installed packages, container images and running services.
2. For each match, note the installed version and whether it is inside the affected range.
3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
```

## FAQ

### Is CVE-2014-5432 being exploited?

There is no confirmed exploitation: CVE-2014-5432 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 2.57% chance of exploitation in the next 30 days (as of 2026-11-26).

### How serious is CVE-2014-5432?

Critical: CVSS 9.8 (v3.0, scored by NVD). Remote, no login needed.

### Which versions are affected by CVE-2014-5432?

Baxter SIGMA Spectrum Infusion System 6.05 (model 35700BAX) with wireless battery module (WBM) version 16..

### Is there a fix for CVE-2014-5432?

No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.

### How do I check if I'm affected by CVE-2014-5432?

Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.

## References

- other: <https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01>

## Attribution

- CVE Program: CVE® is a registered trademark of The MITRE Corporation. CVE records © 1999–2026 The MITRE Corporation. <https://www.cve.org/Legal/TermsOfUse>
- NVD: This product uses data from the NVD API but is not endorsed or certified by the NVD. <https://nvd.nist.gov/vuln/detail/CVE-2014-5432>

HTML: <https://flawdb.com/CVE-2014-5432> · JSON: <https://flawdb.com/api/v1/entries/CVE-2014-5432>