CVE-2014-5431
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 conta…
Mitigation
- No fix listed yet: follow the vendor advisory.
Who should check
- Anyone running Baxter SIGMA Spectrum Infusion System 6.05 (model 35700BAX) with wireless battery module (WBM) version 16
Vulnerability facts
- Affected
- Baxter SIGMA Spectrum Infusion System 6.05 (model 35700BAX) with wireless battery module (WBM) version 16
- Fixed in
- No fix listed yet
- Exploited in the wild
- Not in CISA KEV
- Public exploit / PoC
- None known
- Remote exploitation
- Not remotely exploitable (local or adjacent access needed)
- CVSS
- 6.8 (v3.0, NVD)CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:HOpen in CVSS calculator
Show all 8 CVSS metrics
- Attack vector: Physical
- Attack complexity: Low
- Privileges required: None
- User interaction: None
- Scope: Unchanged
- Confidentiality impact: High
- Integrity impact: High
- Availability impact: High
- EPSS
- 0.38% chance of exploitation in 30 days (percentile 24, 2026-11-26)
- CWE
- CWE-259
- Modified
Description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
Affected and fixed versions
| Package / product | Affected | Fixed | Source |
|---|---|---|---|
| Baxter SIGMA Spectrum Infusion System | 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 | — | cvelist |
Check your systems with an agent
Paste this into Claude Code, Cursor or another coding agent. It only reads; it changes nothing.
Check whether anything I run is affected by CVE-2014-5431. Work read-only: do not install, upgrade, delete or change anything. Details: https://flawdb.com/CVE-2014-5431.md Affected: Baxter SIGMA Spectrum Infusion System 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 Fixed in: no fixed version listed yet Steps: 1. Search this machine and project for Baxter SIGMA Spectrum Infusion System: dependency manifests and lockfiles, installed packages, container images and running services. 2. For each match, note the installed version and whether it is inside the affected range. 3. Report a table: location, installed version, affected (yes / no / unsure), evidence. If affected, say which fixed version or mitigation from the details link applies, but don't apply it.
References
FAQ
Is CVE-2014-5431 being exploited?
There is no confirmed exploitation: CVE-2014-5431 is not in the CISA Known Exploited Vulnerabilities catalog. No public exploit code is known. EPSS estimates a 0.38% chance of exploitation in the next 30 days (as of 2026-11-26).
How serious is CVE-2014-5431?
Medium: CVSS 6.8 (v3.0, scored by NVD). Not remotely exploitable (local or adjacent access needed).
Which versions are affected by CVE-2014-5431?
Baxter SIGMA Spectrum Infusion System 6.05 (model 35700BAX) with wireless battery module (WBM) version 16.
Is there a fix for CVE-2014-5431?
No fixed version is listed in the sources yet. No fix listed yet: follow the vendor advisory.
How do I check if I'm affected by CVE-2014-5431?
Find where you use the affected software (dependency lockfiles, installed packages, container images, servers) and compare the installed version with the affected range above. The agent prompt on this page does this read-only.